4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
z
Speed foot control
Used to continuously adjust the instrument up to maximally the set speed
and intensity values depending on the setting the foot control pedal.
9
The
Sub-screen
of the removed instrument is displayed on the
touchscreen.
¾
Touch the
Direct Starter/speed foot control
key.
ª
If the key is highlighted gray, the direct starter function is switched on.
If the key is highlighted orange, the speed foot control function is switched
on.
4.5.8.7
Setting the spray amount
The spray amount is preset at the factory. However, it can be adjusted using
the control valve at the bottom of the dentist element. This setting is then valid
for all burr drives.
1.
Loosen the ring
(A)
counterclockwise.
2.
Set the spray amount by turning the screw
(B)
.
3.
Check the set spray amount with a burr drive and correct the setting if
necessary.
4.
Tighten ring
(A)
again.
4.5.8.8
Preparing for use of NaCl saline solution
A sterile saline solution instead of spray water is prepared for cooling via the
peristalic pump.
The peristaltic pump hose is a disposable item. To reorder it, see "Spare parts
and consumables" [
.
+