4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
4.5.9.4
Implantology/endodontics treatments with motor
If your treatment center is equipped with the Implantology/endodontics
software option, please observe the instructions provided in the Chapter
"Implantological and endodontic treatments" [
. If you do not have this
option, please note that you have no possibility for setting the electronic
torque.
Endodontics with motor screen
4.5.10
Turbine
The turbine hose is equipped with a standardized coupling according to ISO
9168.
The coolant, i.e. spray or air, and the instrument light can be set in the turbine
sub-screen; see "General instrument functions" [
4.5.11
SPRAYVIT L multifunctional syringe
The multifunctional syringe is used to clean the treatment areas and blow
them dry. It supplies preheated air and water media. The heating cartridge is
located in the handpiece.
4.5.11.1
Safety information
The SPRAYVIT is equipped with extensive safety monitoring functions.
However, please observe the following information.
NOTE: Audible warning signal during CCW rotation
After you start the electric motor with the foot control, an audible warning
signal sounds 6 times if counterclockwise rotation is activated.
WARNING: No electronic torque limitation is available in the motor
screen.
Endodontic files can easily break during operation without torque limitation.
¾
Never perform endodontic treatments without torque limitation. Use an
endodontic handpiece with mechanical torque limitation, e.g. SIRONiTi
from Sirona.
NOTE:
Also observe the operating instructions for the different turbines.
NOTE:
Also observe the operating instructions of the SPRAYVIT L.