61 93 556 D3509
D3509
.
201.01.02
.
02
19.09.2008
84
Sirona Dental Systems GmbH
4 Operation
Operating Instructions TENEO
Dentist element
Explanation of the symbols and labels used
4.5.13.7
Safety checks
In Germany, medical devices are subject to the provisions of the Ordinance
on the Installation, Operation and Use of Medical Devices (Medizinprodukte-
Betreiberverordnung – MPBetreibV). Safety checks must be performed and a
medical product log must be kept.
For more information, refer to "Safety tests for systems with HF surgical
equipment" [
.
4.5.14
Implantological and endodontic treatments
The Treatment functions support implantological and endodontic treatments.
The speed and torque of the instrument tip can be precisely adjusted and
saved for later reference if desired. This is possible for every work step for
implantology treatments. A selection of the most popular file systems with the
speed and torque values recommended by the manufacturer is available for
endodontic treatments. The file assortment and sequence can be set by the
user.
The treatment center allows for the management of two implantological and
up to five endodontic treatments with individual settings for each user profile.
The therapy functions can only be used in conjunction with the precisely
adjustable brushless motors. If the adjacent display appears when the
Treatment function is switched on (see below), the software has detected that
the burr drives marked with a warning triangle are not suitable for treatment
purposes. In this case, please contact your dental depot.
In order to supply the burr drive with a sterile saline solution during
implantological treatments, a peristaltic pump must be attached to the dentist
element; see "Preparing for use of NaCl saline solution" [
Neutral electrode connection
Neutral electrode with high frequency
referenced to earth
SIROTOM HF electrosurgical handpiece, type
BF applied part
Intermittent operation 25%
Duty time 10s
Cycle time 40s
HF frequency 1MHz
Symbol for non-ionizing radiation
DAB 25% ED 10s SD 40s
1 MHz / 50 W
300 - 600 Ohm