4 Operation
Sirona Dental Systems GmbH
Dentist element
Operating Instructions TENEO
¾
Use the – and + keys to set the number of work steps.
ª
The number of work steps is displayed on the left side:
Implant 1... X.
4.5.14.2.9
Preparing the treatment center for sterile operation
The requirements for sterile operation must be fulfilled for surgical
interventions. It is recommended to work exclusively with the cursor control to
avoid touching the user interface.
Covering the dentist element with a sterile drape
The dentist element can be covered with a sterile drape for sterile operation.
A rectangle must be cut out of the drape to allow for operation and visibility of
the EasyTouch.
Use of separate motor holder
Since the instrument holder of the dentist element is inaccessible due to the
sterile drape, the burr drive being used can be deposited in a separate motor
holder. This separate holder can be placed on top of the dentist element.
The motor holder can be sterilized.
To reorder the motor holder, see "Spare parts and consumables" [
.
The Implant motor becomes the active instrument as soon as the hose of the
implantology instrument is connected to the dentist element.
Attaching the instrument hose cover
The instrument hose of the burr drive used can be covered with a sterile paper
sleeve. The instrument hose sleeves can be ordered from a specialized
dealer.
4.5.14.3
Endodontics
4.5.14.3.1
File selection
In the Simple Endodontics program
A list of the files available for this endodontic treatment is displayed over the
full height of the touchscreen in the
Simple Endodontics progr
am
. Seven files
can thus be displayed simultaneously.
The speed and torque values preset or set for the selected files are
automatically used.
9
An endodontic treatment is selected.
9
The
Simple Endodontics program
is displayed on the touchscreen.
¾
Use the
↑
and
↓
keys to select the file that you would like to use.
ª
The selected file is highlighted orange.
In the Advanced Endodontics program
A maximum of three files is displayed in the
Advanced Endodontics program
.
Speed and torque values can be modified individually, see "Setting the speed
and torque" [
94]
.
9
The
Simple Endodontics program
is displayed on the touchscreen.