Use of the X-ray sensor
Sirona Dental Systems GmbH
Hygienic protection
Operating Instructions and Installation
XIOS USB
8.2
Hygienic protection
Hygienic protection
Please use Sirona hygienic protective sleeves for hygienic protection.
z
Prior to each application on a new patient, the sensor must be covered
with a new hygienic protective sleeve.
z
Do not use hygienic protective sleeves more than once!
z
Do not bend the sensor cable when sliding the hygienic protective sleeve
over it!
z
Please use the hygienic protective sleeve which correctly matches the
sensor.
- Hygienic protective sleeve for sensor size 1
300 pcs, Order No.: 61 76 577
- Hygienic protective sleeve for sensor size 2
300 pcs, Order No.: 61 76 569
8.2.1
Sliding the hygienic protective sleeve over the
sensor
Handling
1.
Please select a hygienic protective sleeve with a size that matches the
sensor.
2.
Slide the sensor into the hygienic protective sleeve.
3.
Position the sensor holder on the hygienic protective sleeve.
Handling information for the parallel technique
If the supplied XIOS sensor holder set is used, hygienic protection is provided
between the sensor and the holder. The sensor must not be glued onto the
unprotected sensor holder.
CAUTION:
Use only the XIOS hygienic protective sleeves offered by Sirona, especially
in connection with the XIOS sensor holders.
NOTE:
Make sure that the cable connecting the sensor to the USB module is run out
of the patient's mouth in such a way that the patient cannot bite it.
NOTE: Size of the hygienic protective sleeve
The hygienic protective sleeve is slightly undersized to prevent the sensor
from slipping.
NOTE:
The exact position depends on the sensor holder and the X-ray exposure
area. The following sections contain specific information on how to position
different sensor holders.
CAUTION:
Never slide the hygienic sleeve over the entire holder with sensor!