Use of the X-ray sensor
Operating Instructions and Installation
XIOS USB
Parallel technique with radiation limiter!
5.
For the right upper jaw and left lower jaw:
Position the sensor holder
in the middle of the sensor as shown in the illustration. The edge of the
sensor holder must be flush with the edge of the sensor.
X-ray exposure
1.
Position the sensor in the patient’s mouth.
2.
Place the X-ray tube assembly in the correct position. Change the
position of the sensor holder if necessary.
3.
Release an X-ray exposure.
4.
Discard the sensor holder tab and the hygienic protective sleeve
following the patient examination.
5.
The guide rod and localizer ring must be sterilized.
8.3.3
Bite wing exposures
Explanation
The "bite tab" type sensor holder is available for bite wing exposures.
z
This sensor holder and the matching localizer ring are color-coded
red
.
z
The straight guide rod and red localizer ring for bite wing exposures must
be used.
z
The following illustrations show how to attach the sensor holder to the
hygienic protective sleeve with sensor.
Preparation
1.
Plug together the straight guide rod
(C)
and the red localizer ring (
B)
of the
XIOS holder system for
bite wing exposures
.
2.
Select the (bite tab type) red sensor holder for bite wing exposures and
plug it onto the guide rod
(C)
.
3.
Slide the sensor into the hygienic protective sleeve. When doing so,
follow the instructions for sensors.