Instruction Manual of 

Waterproof Digital Thermometer Model PC-9230 / 9235 

These products are designed to measure temperature. Do not use them for other 
purposes. Read this manual thoroughly before use, and keep it for future reference. 

 

Features 

 

 

 

Fast response

 achieved by ultra-narrow sensor needle with a diameter of 1.5 mm 

Display orientation can be right side up / left side up 
Maximum / minimum temperature shown by a single press 

Waterproof

 body resists water splashes 

 

Important Notices 

 

 

 

 

Beware of Explosion 

PC-9230 and PC-9235 are not explosion-proof. Never use them in an atmosphere 
containing flammable gases. 

 

 

Cautions 

To use the thermometer properly, follow the instructions below. 

 

Do not use the device as a clinical thermometer. 

 

Hold the device in your hand while in use.

 Leaving the device in a hot pot or the 

like may result in a deformation or malfunction.

 

 

Do not use the sensor sheath as a handgrip,

 or the device may 

fall and get broken.

 

Sensor  sheaths  are  made  only  to 
protect sensor needles when not in 
use. 

 

 

Do not drop or give a shock to the device, as it is a sensitive instrument. 

 

Do not use the device under direct sunlight or aside a heat source, or it may worsen 
the measurement accuracy and result in a deformation or malfunction. 

 

Do not use the device in an environment with electrical noise, or the display may 
become  unstable  or  the  measurement  error  may  increase.  Some  induction 
cooktops may affect the measurement, although the device is designed to resist 
them.

 

 

Do  not  measure  temperature  out  of  the  measurable  range,  or  it  may  result  in 
malfunctions.

 

 

To eliminate the effect from ambient temperature and improve accuracy, insert the 
needle in a target for 20 times as long as the diameter of the sensor needle, thus 
30 mm.

 

 

Do not remain the needle inserted in a target for more than a minute if it is hot, or 
the heat may reach the main body and lead to a measurement error or deformation. 

 

For  your  reference,  PC-9230  and  PC-9235  needs  only  10  seconds  to  finish 
measurement when the needle in a room temperature is inserted in boiling water.

 

 

Stay distant from the target and be sure not to burn yourself when it is hot. 

 

After  measuring  high  temperature,  wait  until  the  needle  cools  down  before 
installing the sensor sheath. 

 

For PC-9230, round holes on the sensor sheath faces up when it covers the needle. 
For PC-9235, a projection on the sensor sheath faces up when it does the same. 

 

Do not leave the device in a place like an automobile on a fine summer day, or the 
extreme heat may damage the thermometer. 

 

Remove batteries if the device is not used for a long period of time. Otherwise, the 
batteries may leak resulting in malfunctions.

 

 

Wet hands or water splashes do not affect the function of the device. However, 
underwater use is forbidden. 

 

Do not disassemble or customize the device, or it may cause a fatal malfunction.

 

 

PC-9230  and  PC-9235  can  be  fully  rinsed  with  tap  water  unless  they  are  kept 
submerged in water. After washing, wipe out with a dry cloth thoroughly. 

 

Do not wash or wipe the device with alcohol, thinner, or other organic solvents. Do 
not throw it in a dryer, either.

 

 

For repair or calibration, contact us or a retailer from which you have purchased.

 

 

 