cassette. Fit the retaining ring over the sampler top and screw down firmly to secure 
the cassette within the sampler and seal the cassette inlet and outlet.
Ensure that the grit pot is securely fitted over the ridge around the bottom of the 
sampler body.

Flow Calibration

Connect the inlet hosetail of the sample pump to the air outlet of the cyclone sampler 
with a length of flexible tubing. Connect the flow calibration device (such as a chek-
mate flowmeter or rotameter) to the air inlet of the cyclone sampler using a second 
length of flexible tubing.
Calibrate the sample flow rate to 3.0 l/min.
Disconnect the tubing from the cyclone sampler inlet and outlet. Remove the cassette 
from the cyclone sampler.
Hold the cassette upside down and prise the two halves apart using the side tab. 
Remove and discard the calibration filter. Place the pre-weighed filter to be used for 
the sample, onto the filter support grid and re-assemble the cassette. Fit the cassette 
into the cyclone sampler.
Quickly re-check the flow rate and make any final adjustments. Switch the sample 
pump off.
The cyclone sampler is now ready for sampling.

After Sampling

Away from the sample location quickly check the sample flow rate, then disconnect 
the tubing from the cyclone sampler air inlet and outlet. 
Carefully remove the cassette from the cyclone sampler and split the two parts of the 
cassette apart.

Post Weighing the Filter

Carefully lift the filter out of the cassette using tweezers. Post weigh the filter.
The difference between the pre and post weights will give the respirable sample.

Transportation

If further analysis is required, prepare the sample for transportation to the laboratory by 
carefully placing the filter back in the cassette, and fitting the yellow, U shaped sealing 
clip over the cassette inlet and outlet holes. Use of the sealing clip prevents unwanted 
ingress or loss of contaminant during transport. Alternatively, a filter transport cassette 
(SKC P/N 225-67-10) can be used, freeing the filter cassette for immediate re-use.

Summary of Contents for 225-69

Page 1: ...etal filter support grid and a sealing clip to prevent contamination of the sample during transport for laboratory analysis For detailed instructions on air sampling in general and use of the cyclone sampler in particular refer to SKC s 224 G1 Step by Step Guide to Air Sampling It is also recommended to obtain and read the document MDHS 14 4 General methods for sampling and gravimetric analysis of...

Page 2: ...fit Fitting the cassette into the cyclone sampler Referring to Figure 2 ensure that the large O ring is fitted to the sampler body beneath the sampler insert and the small O ring is fitted over the short tube in the middle of the insert Fit the assembled cassette into the cyclone insert with the cassette top upwards Check that the thin O ring is fitted into the recess in the underside of the sampl...

Page 3: ... onto the filter support grid and re assemble the cassette Fit the cassette into the cyclone sampler Quickly re check the flow rate and make any final adjustments Switch the sample pump off The cyclone sampler is now ready for sampling After Sampling Away from the sample location quickly check the sample flow rate then disconnect the tubing from the cyclone sampler air inlet and outlet Carefully r...

Page 4: ...ptember 2018 SKC Limited 11 Sunrise Park Higher Shaftesbury Road Blandford Forum Dorset DT11 8ST United Kingdom T 44 0 1258 480188 F 44 0 1258 480184 E info skcltd com W www skcltd com A member of the SKC global group of companies ...

Reviews: