Battery and Power Supply Specifics
Charging the Batteries
8 - 2
Strobe Light
User
Manual
Low Battery Icon
- Outline blinking (very little time
left)
The strobe has a protection feature that prevents the
strobe from operating if the battery voltage is too low.
This condition is indicated by no flash and the display
shows
LO BAT
. At this time the batteries must be
recharged (see
Charging the Batteries
section below)
or powered by the power supply / charger (see
External Power Supply / Charger
section below).
Charging the Batteries
The unit may be charged at any time. You do not
need to wait until the low battery condition is
indicated.
To charge the Strobe Light with the power supply /
charger:
•
Release the trigger so the strobe is off.
•
Connect the charging cable into the charging
socket (located below the display panel behind
the handle).
•
Connect the charger into an AC mains wall
outlet (115 / 230 Vac).
CAUTION:
Use of chargers other than the one supplied will
damage the stroboscope and void the warranty.
When charging, the strobe will indicate
CHRGE
in the
bottom right of the display. The charger will fast
charge the batteries for about 4 - 5 hours and then
trickle charge the batteries.
Allow the charger to charge the batteries until the
display shows
DONE
for peak battery life