The ROUTE Module
Viewing Dynamic Measurement Data
Figure 3 - 6.
A Spectrum Display Screen.
overall value
POINT info area; machine name, POINT ID, POINT description, alarm condition
cursor
current cursor frequency and amplitude
alarm indicators
previous measurement indicator
% difference from previous measurement
set cursor frequency as running speed
retake meas.
view / add coded and type-in notes
escape without storing data
Items on the spectrum display screen include:
Overall value
– The measurement’s overall value displays in the display’s upper left
corner.
Cursor position/amplitude
– Use the left/right arrow buttons to move the spectrum’s
cursor. The cursor position and amplitude display in the upper right corner.
Percent Changed Bar
- The bar to the right of the spectrum provides information
about the overall value of the measurement in comparison to alert and danger alarms
(yellow and red arrows, respectively) and to the previous measurement (green arrow).
The current measurement value is represented by the colored bar, which displays red if
the POINT is in alarm, yellow if it is in alert, and green if there is no alarm. If the
measurement amplitude extends beyond the display, an “overall overflow” indicator
appears at the top of the bar.
SKF Microlog - GX Series
3 - 11
User Manual