The Analyzer Module
How to Collect Standard Vibration Measureme
nts
To expand the measurement parameter list:
•
Press the Shift / up arrow button to display additional function buttons, then press
the
Expand
function button. The full measurement parameter list displays.
Figure 4 - 7.
The
Acceleration
Measurement’s Setup Screen (Expanded View).
To alter the measurement’s setup and then perform the measurement:
•
Use the up / down arrow keys to highlight the measurement parameter to change,
then use the right arrow to display the parameter’s options.
•
Highlight or enter the desired option, then press the left arrow key (or Enter button)
to select the highlighted option.
Analyzer Measurement Setup Fields
If you wish to alter a standard vibration measurement’s default settings, enter
appropriate setup information in the following fields:
Num Channels –
Choose from
CH1, CH1 & CH2, CH1 & CH2 & CH3,
or
CH1 & CH2 &
CH3 & CH4.
A channel is a single point of monitoring, including the sensor and the
hardware needed to condition the input signal and to compare the input signal to pre-
determined setpoints.
For measurements taken with a Triax sensor, select
CH1 & CH2
& CH3
.
Sensor –
Select the appropriate preconfigured sensor from the setup screen’s context
sensitive sensor list.
Note that the specified sensor type determines available options
and engineering units for subsequent setup fields. Also note that in Reduced view setup
screens, the
Sensor Units
,
Sensitivity
, and
ICP Supply
parameters only display if the
Variable
sensor is selected.
SKF Microlog - GX Series
4 - 9
User Manual