Error Message
MESSAGE
WHAT IT MEANS
ACTION
Appears when the
batteries can not provide
enough power for a test.
Replace the batteries immediately.
Appears when inserting a
used test strip.
Test with a new test strip.
System operation range is 10~40°C
(50~104°F). Repeat the test after the
meter and test strip have reached
the above temperature.
Re-test with a new test strip.
Review the instructions and re-test
with a new test strip. If the above
steps do not work, please contact
your local Customer Care Line at
for help.
Appears when ambient
temperature is below
system operation range.
Appears when ambient
temperature is above
system operation range.
Remove the strip after
applying blood to the
absorbent hole.
Problem with the meter.