Page 8
6701 OPERATOR • REV2
WARNING
The antistatic properties of the table are
dependent on the use of the original pad
set which was furnished with the table or
an alternate approved replacement.
WARNING
Certain accessories may limit weight
capacities. Check with your SKYTRON
representative.
WARNING
SKYTRON Products are guaranteed
for proper performance with the use of
genuine SKYTRON accessories.
Accessories and products not furnished
by SKYTRON have not been tested for
proper performance and safety. Such
applications or use are at the discretion
of the user to ensure patient and staff
safety.
CAUTION
Do not use the table if any of the
inspection points fail.
WARNING
Personal injury to patient or staff may
result from a lack of proper maintenance
of this equipment.
NOTE
Normal table top position is with the
head (and back) section over the power
cord end of the base.
WARNING
Always lock the table top in position after
rotation. DO NOT rotate the top with
an unevenly distributed patient weight
load as instability may result.
WARNING
•Make sure the TOP ROTATION LOCK
HANDLE is tightened and the brakes are
set before transferring the patient.
•Exercise caution with the table top
rotated 90° to the base since an
improperly distributed patient load
may cause the table to be tipped over.
A table support rod is required for 90°
positioning. See Specialty Positioning.
WARNING
Consult manufacturer's instructions
when using high frequency surgical
equipment, cardiac defibrillator and
cardiac defibrillator monitors.
WARNING
When an antistatic pathway is required,
the table has to be used on an antistatic
floor.
SPECIAL USER ATTENTION