Page 13
6701 OPERATOR • REV2
2-3. Battery Operation
a.
Make sure the Battery Indicator and Main
Power Switch indicator lights, on the electrical
panel, are OFF. See figure 2-2. If the indicator
lights are ON, turn AC120V operation OFF with
the main power switch.
NOTE
The table will operate correctly on battery
power with the power cord connected
to a wall outlet or disconnected.
b.
Press the BATT button on the hand-held
pendant control. The pendant control buttons
and the red BATTERY indicator light, located in
the upper right corner of the pendant control, will
illuminate.
c.
The table is now ready for BATTERY
operation.
d.
To extend the battery charge life, turn the
BATTERY power OFF with the pendant control
when the table is not going to be used.
NOTE
Battery Operation must be turned
OFF at the pendant control. It cannot
be turned OFF using the main power
switch.
2-4. Automatic Shut-Off
a.
To prevent unnecessary discharge of the
battery, a timer is built into the battery circuit. This
timer will automatically shut the battery power OFF
after 1½ hours of table inactivity.
b.
To turn the table ON again, press the BATT
button on the pendant control, the pendant control
buttons and the red indicator light will illuminate.
NOTE
Turning the Main Power Switch ON will
change the table operation to 120 VAC
power.