Page 19
6701 OPERATOR • REV2
Figure 2-12. Kidney Lift Positioning
NOTE
To prevent damage to the kidney lift,
a safety interlock prevents the kidney
lift from going up if the back section is
raised above horizontal. An audible
alarm will sound .
i. Return To Level.
To return the table top to
a level position, press the LEVEL button (figure
2-13).
NOTE
Elevation and brake system functions
are not affected by the return to level
function.
Figure 2-14. Beach Chair Positioning
j. Beach Chair.
To place the top in the beach
chair position from a level position, press the BEACH
CHAIR button (figure 2-14). The back section will
raise, the leg section will lower and the Trendelen-
burg positioning will function simultaneously. The
functions will stop when Trendelenburg reaches it
limit.
h. Kidney Lift.
To raise the built-in kidney lift,
press the KIDNEY UP button (figure 2-12). Up to
6 inches of lift can be achieved. Press the KIDNEY
DOWN button to lower the kidney lift.
Figure 2-13. Return To Level
6"
KIDNEY UP
KIDNEY DOWN
RETURN
TO LEVEL
BEACH
CHAIR