background image

Page 22

6701 OPERATOR • REV2

WARNING

Ensure that the Leg and Back sections  
are properly engaged and secured to 
pins before use to prevent injury.

NOTE

The Leg and Back sections are labeled 
for proper orientation.  The Leg section 
cannot be installed on the Back section 
pins.

2-11. Table Top Rotation.

NOTE

Normal  table  top  position  is  with  the 
head (and back) section over the power 
cord end of the base.

 

  a.

  The  table  top  can  be  horizontally  rotated 

210° without having to rotate the entire table. Refer 
to  figure  2-21.    To  rotate  the  top,  turn  the  TOP 
ROTATION  LOCK  HANDLE  counterclockwise  to 
release, grasp the table by the head end and rotate 
the top counterclockwise.  Lock the top in position 
by tightening the TOP ROTATION LOCK HANDLE 
clockwise.

WARNING

Always lock the table top in position after 
rotation.  DO NOT rotate the top with 
an unevenly distributed patient weight 
load as instability may result.  

Figure 2-21. Top Rotation

2-10.  Leg and Back Section Removal.

 

a.  

The Leg Section and the Back Section on 

the 6701 Hercules tables are both removable.  See 
figure 2-20 (Leg Section shown).

 

b.  

 To remove either section, level the table top, 

simultaneously depress both release levers and pull 
the section out.   Press the LEG DOWN or BACK 
DOWN button on the pendant control to position 
the attachment pins down and out of the way.

NOTE

To  make  the  Back  Section  easier  to 
handle, remove the Head Section and 
X-ray Top prior to removing the Back 
Section.

Figure 2-20.  Leg Section Removal

 

  c. 

 To install the Leg Section, press and hold the 

LEG UP button until the attachment pins completely 
stop. Install the section on the pins.  Level the table 
top and pull out on the section to make sure the 
release levers are completely locked.

  d. 

 To install the Back Section if the attachment 

pins  are  not  aligned,  press  and  hold  the  BACK 
DOWN button until the attachment pins completely 
stop (40° down). Press REV TREND to bring the 
pins up to a level position.

FOOT/LEG SECTION

LOCKING

LEVERS

TOP ROTATION

HANDLE

Summary of Contents for 6701 hercules

Page 1: ...SURGICAL TABLE 6701 HERCULES OPERATORS MANUAL TEC D 0010 REV2 6 11...

Page 2: ......

Page 3: ...1 Electrical Power 12 2 2 AC 120V Operation 12 2 3 Battery Operation 13 2 4 Automatic Shut Off 13 2 5 Charging the Battery 14 2 6 Positioning Functions 15 a Floor Lock Brake system 15 b Trendelenburg...

Page 4: ...EXERCISE CAUTION WITH THE TABLE TOP ROTATED 90 TO THE BASE SINCE AN IMPROPERLY DISTRIBUTED PATIENT LOAD MAY CAUSE THE TABLE TO BE TIPPED OVER TABLE TOP WARNING Table Capacity Lift 1200 lbs Articulate...

Page 5: ...ector removable 12 1 2 19 1 4 23 24 1 4 14 1 2 19 3 4 6 3 79 TOP VIEW 90 60 14 1 2 37 SIDE VIEW 8 1 2 5 3 4 19 22 41 MAX 23 MIN END VIEW UNIT TO BE USED ONLY IN SPECIFIED ENVIRONMENTAL CONDITIONS TEMP...

Page 6: ...for all personnel that may operate this table The maximum lifting capacity of the 6701 Herculestableis1 200poundsandthemaximum articulation weight capacity is 1 000 pounds When lifting or articulatin...

Page 7: ...connected to a wall outlet or disconnected NOTE Battery Operation must be turned OFF at the pendant control It can not be turned Off using the main power switch NOTE Turning the Main Power Switch ON w...

Page 8: ...o be unlocked the patient must be centered and evenly distributed on the table top i e supine or prone position withthetableloweredtoitslowestheight position The maximum patient weight shouldnotexceed...

Page 9: ...for proper orientation The Leg section cannot be installed on the Back section pins WARNING The Leg section may hit the table base or the floor if both the leg and elevation systems are placed in the...

Page 10: ...fromalackofpropermaintenance of this equipment NOTE Normal table top position is with the head and back section over the power cord end of the base WARNING Alwayslockthetabletopinpositionafter rotatio...

Page 11: ...cluding gloves masks and eye protection when cleaning the surgical table CAUTION Thoroughly read and follow the manu facturer s directions for all cleaning fluids DONOTusecleanerscontaining phenolics...

Page 12: ...lex kidney lift return to level beach chair and the floor lock brake system Manual controls are provided for head section positioning table top rotation emergency brake release back section removal an...

Page 13: ...on Buttons 1 4 Floor Lock Brake System The floor lock brake system consists of four self leveling hydraulic brake cylinders which raise and support the table base off from the casters Press the TABLE...

Page 14: ...securelyattached to the table To install the power cord align the cord connector with the base connector insert the cord and twist clockwise to lock the cord into the connector See figure 2 1 Plug the...

Page 15: ...ill illuminate c The table is now ready for BATTERY operation d To extend the battery charge life turn the BATTERY power OFF with the pendant control when the table is not going to be used NOTE Batter...

Page 16: ...e charge level turn off and light in sequence again NOTE When the red light starts to blink indicating low power in battery the table will operate for approximately 5 continuous minutes typically long...

Page 17: ...op i e supine or prone position withthetableloweredtoitslowestheight position The maximum patient weight shouldnotexceed500pounds Tabletop rotation must be in normal orientation that is the back secti...

Page 18: ...imize patient safety utilize properrestraintmethodsduringextreme Trendelenburg positioning c Lateral Tilt To achieve lateral tilt right as viewed from the head end of the table press the TILT RIGHT bu...

Page 19: ...t is not all the way down If the kidney lift is raised an audible alarm will sound when the BACK UP button is pressed NOTE Ifbacksectionisbelowhorizontal BACK UP function is limited to 800 pound patie...

Page 20: ...if both the leg and elevation systems are placed in their full down position g Flex Positioning To place the table top in a flex position from horizontal press the FLEX button figure 2 11 To return t...

Page 21: ...to level function Figure 2 14 Beach Chair Positioning j Beach Chair To place the top in the beach chairpositionfromalevelposition presstheBEACH CHAIR button figure 2 14 The back section will raise the...

Page 22: ...In case of a power failure or an electrical problem within the table the emergency brake release system can be used to move the table The control knob for this function is located on the side of the t...

Page 23: ...ve has been operated the BRAKE UNLOCKbuttononthependantcontrol will have to be pressed before brakes will lock again 2 9 Head Section a A quick release positioning bar located under and to the front o...

Page 24: ...ity may result Figure 2 21 Top Rotation 2 10 Leg and Back Section Removal a The Leg Section and the Back Section on the 6701 Hercules tables are both removable See figure 2 20 Leg Section shown b To r...

Page 25: ...te approved replacement 2 12 Positioning The use of certain optional accessories available from SKYTRON further extend the positioning capabilities of the 6701 Hercules Tables Refer to the following P...

Page 26: ...Page 24 6701 OPERATOR REV2 6701...

Page 27: ...Page 25 6701 OPERATOR REV2...

Page 28: ...and adjust if necessary 3 Pressurecheck withagauge thepressure relief valve 4 Check all mechanical adjustments and adjust as necessary 5 Check hydraulic fluid level 6 Lubricate the slider assembly 7 C...

Page 29: ...luids DONOTusecleanerscontaining phenolics Apply cleaning fluid liberally to top and sides of each pad and wipe with a clean lint free cloth Using a clean damp lint free cloth wipe the pads to remove...

Page 30: ...presentative To obtain SKYTRON authorized service or preventive maintenance contracts contact your nearest SKYTRON representative listed below Or contact SKYTRON 5085 Corporate Exchange Blvd S E Grand...