Page 26
6701 OPERATOR • REV2
3-2. Preventive Maintenance
The following preventive maintenance checks and
services are recommended to ensure the service-
ability and proper operation of your SKYTRON
Surgical Table, and should only be performed by
qualified SKYTRON trained personnel.
a.
During normal cleaning, a general visual
examination should be made checking for leaks,
loose bolts or parts, and cracked, chipped, or
missing paint. Any necessary repairs should be
made.
b.
Annually or as required based on usage,
the following checks and services should be
performed:
1. Check all hydraulic fittings, mini-valves and
slave cylinders for proper operation and any
signs of leaks.
2. Check the hydraulic speed controls and
adjust if necessary.
3. Pressure check (with a gauge) the pressure
relief valve.
4. Check all mechanical adjustments and
adjust as necessary.
5. Check hydraulic fluid level.
6. Lubricate the slider assembly.
7. Check function of foot leg release levers,
(release knob early models). Lubricate as
necessary.
3-1. Routine Inspections
The following inspections should be done before
and after each use of the table.
a.
Inspect all table pads for damage.
b.
Inspect all table top sections for damage.
c.
Inspect the table top assembly, all top sections
and the base for stability.
d.
Inspect the power cord and plug for any signs
of burns or damage.
e.
Test the operation of the main power switch.
f.
Test all functions of the pendant control for
proper table movement.
g.
Inspect the table base surface and the floor
for any signs of oil leaking.
CAUTION
Do not use the table if any of the
inspection points fail.
SECTION III MAINTENANCE