Page 27
6701 OPERATOR • REV2
3-3. Cleaning Recommendations
WARNING
Personal injury to patient or staff may
result from a lack of proper maintenance
of this equipment.
NOTE
Always follow current AORN Journal
Guidelines to ensure proper cleaning
and disinfection procedure.
CAUTION
Caution should be taken when cleaning
the table to prevent excessive fluid entry
into electrical connectors.
The following procedures should be followed when
cleaning the surgical table between cases.
Place table top in level position prior to starting
cleaning procedure.
WARNING
Always follow OSHA blood-borne
pathogens standards for protective
clothing, including gloves, masks and
eye protection when cleaning the
surgical table.
Remove major contaminants from the table
with disposable materials following appropriate
biohazard waste disposal procedures.
Remove all table pads and place them on a flat
surface for cleaning.
CAUTION
Thoroughly read and follow the manu-
facturer's directions for all cleaning
fluids. DO NOT use cleaners containing
phenolics.
Apply cleaning fluid liberally to top and sides of
each pad and wipe with a clean lint-free cloth.
Using a clean, damp, lint-free cloth, wipe the pads
to remove the cleaning fluid.
Using a clean, dry, lint-free cloth, wipe the pads to
remove all moisture.
Repeat the steps to clean the bottom of the each
pad.
CAUTION
When using spray cleaners DO NOT
spray fluids directly into electrical
receptacles or micro switches.
Repeat cleaning procedure for all table surfaces
including the top, sides, elevation column, base
and all accessories.
CAUTION
Before replacing pads on the table,
make sure the pads and all mating
surfaces are completely dry. Moisture
trapped between the pads and mating
surfaces may cause distortion of table
tops.
When the cleaning procedure is complete, replace
all pads and accessories as applicable.
Remove pendant control from table side rail and
apply cleaning solution to the pendant control and
cord.
Use a clean cloth dampened with water to remove
cleaning solution.
Use another clean damp cloth to remove any
remaining residue.
Install pendant control on side rail for storage when
cleaning procedure is complete.