Page 6
6701 OPERATOR • REV2
WARNING
To maximize patient safety, utilize
proper restraint methods during extreme
Trendelenburg positioning.
NOTE
To prevent table damage, a safety
interlock limits lateral tilt positioning to
20° if Trendelenburg positioning exceeds
20°. An audible alarm will sound.
WARNING
To maximize patient safety, utilize proper
restraint methods during extreme lateral
tilt positioning.
NOTE
To prevent damage to the kidney lift,
a safety interlock prevents the back
section from going above horizontal if
the kidney lift is not all the way down.
If the kidney lift is raised, an audible
alarm will sound when the BACK UP
button is pressed.
NOTE
If back section is below horizontal, BACK
UP function is limited to 800 pound
patient weight.
WARNING
•Do Not unlock brakes when patient is
on the table. An uneven patient weight
load may cause instability.
•If circumstances demand table brakes
to be unlocked, the patient must be
centered and evenly distributed on the
table top (i.e. supine or prone position)
with the table lowered to its lowest height
position. The maximum patient weight
should not exceed 500 pounds. Table top
rotation must be in normal orientation,
that is, the back section over the long
end of the table. Patient's head must be
on the head section. Head section must
be attached in its normal orientation to
the table's back section.
•Prior to unlocking brakes, check for
obstructions on the floor that might
prevent the table from moving smoothly
to new location. Relock the brakes
immediately once the final position
is reached and before commencing
surgery. Table brakes should remain
locked at all times if patient weight
exceeds 500 pounds.
NOTE
With an evenly distributed patient weight
load, all table positioning functions will
operate smoothly and quietly with a
patient weight of up to 1,000 pounds.
NOTE
To prevent table damage, a safety
interlock limits Trendelenburg positioning
to 20° if lateral tilt positioning exceeds
20°. An audible alarm will sound.
SPECIAL USER ATTENTION