Page 7
6701 OPERATOR • REV2
NOTE
The emergency back-up control
switches will function when the table is
operating on 120VAC power, battery
power, or turned off.
NOTE
•The Emergency Brake Release
Valve must be closed and tightened
(counter-clockwise) before activating
any hydraulic function.
•If the Emergency Brake Release
Valve has been operated, the BRAKE
UNLOCK button on the pendant control
will have to be pressed before brakes
will lock again.
NOTE
To make the Back Section easier to
handle, remove the Head Section and
X-ray Top prior to removing the Back
Section.
WARNING
Ensure that the Leg and Back sections
are properly engaged and secured to
pins before use to prevent injury.
NOTE
The Leg and Back sections are labeled
for proper orientation. The Leg section
cannot be installed on the Back section
pins.
WARNING
The Leg section may hit the table base
or the floor if both the leg and elevation
systems are placed in their full down
position.
NOTE
When REFLEX button is activated, if
Kidney Bridge is up, the back section
will not go above horizontal.
NOTE
To prevent damage to the kidney lift,
a safety interlock prevents the kidney
lift from going up if the back section
is raised above horizontal. An audible
alarm will sound.
NOTE
Elevation and brake system functions
are not affected by the return to level
function.
CAUTION
The Back Section/Kidney Lift safety
interlock is not operational when the
emergency back-up control switches
are used.
CAUTION
The EMERGENCY BRAKE LOCK
switch does not activate the brake
system timer. The switch must be held
until the brakes are completely locked,
approximately 10 seconds.
SPECIAL USER ATTENTION