Page 18
Handle Cleaning & Sterilization
The sterilizable handles are constructed of heat
resistant, impact resistant plastic. They should
be cleaned with mild alkaline cleaning products
WITHOUT active chlorine. Thoroughly rinse off all
cleaners with water. Ensure that the open side of
handles are face down.
Recommended sterilization parameters for steriliz-
able handle:
Steam Sterilization:
134°C (273°F) : 2.3 bars (33.3 PSI)
4 minutes
Always consult current AORN journal recommen-
dations for proper sterilization procedures.
Adjustments
As part of a regular preventive maintenance pro-
gram, it is suggested that a check of the various
positioning axes be made to verify correctness of
tension adjustment. If any lighthead drift is noticed,
all that is usually necessary is a minor adjustment.
Readjustment should be made as per the appro-
priate instructions contained in the Maintenance
and Adjustment Manual. Also, during a scheduled
cleaning of the lighthead interior, lubrication of the
various moving parts is desirable.
MAINTENANCE
General
To insure proper operation and to extend the life
of your SKYTRON surgical lighting system, the
following preventive maintenance procedures are
recommended.
NOTE
All repairs should be made using only
authorized SKYTRON replacement
parts.
Daily Maintenance
Daily or between cases, the lighthead exterior
should be wiped down with a mild cleaning agent
which will not affect the painted or acrylic parts.
Do not apply or spray cleaning agents directly on
the lighthead.
CAUTION
Use of incompatible cleaning agents
will cause damage to the fixture. Avoid
the use of cleaning solutions which
contain high concentrations of alcohol,
ethelene glycol, phenol, iodophors, or
glutaraldehyde based disinfectancts.
Staining, pitting, discoloration and dif-
fuser cracking may occur if these are
used.
•Only dissinfectants and cleaners may be used that
are certified from the manufacturer for compatibility
with the following materials:
-- Polycarbonate (PC)
-- Polyvinylchloride (PVC)
-- Silicones
•Always consult with the manufacturer of the clean
-
ing agent for proper application and use. Always
spot test on an inconspicuous area before use.
•Avoid personal injury. Do not attempt to clean light
-
head unless power is turned off at wall control.
•Avoid using excessive amounts of spray cleaners.
Leakage of fluids into the interior of lighthead may
cause corrosion of electrical components.
Summary of Contents for AURORA LED 5 4000K
Page 1: ...REV 9 08 INSTALLATION INSTRUCTIONS LED SERIES SURGICAL LIGHTS...
Page 18: ...Page 16...
Page 20: ...Page 18...
Page 26: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 27: ...LED SERIES SURGICAL LIGHTS OPERATORS MANUAL TEC B 0002 REV0 1 09...
Page 47: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 48: ...PARTS CATALOG LED SERIES SURGICAL LIGHT TEC B 0009 REV3 3 10...
Page 54: ...Page 5 17 Stand Model Support Post Assembly page 40...
Page 65: ...Page 16 6 LED POD ASSEMBLY 57 010307 03mj 3 2 1 4 10 5 6 7 12 13 15 14 15 13 16 8 11 9...
Page 73: ...Page 24 10 LIGHTHEAD ASSEMBLY 1 2 6 7 8 9 11 13 14 12 10 5 3 4 57 031408 06mj...
Page 81: ...Page 32 13 TV LIGHTHEAD ASSEMBLY 1 3 4 5 6 7 8 9 10 11 12 15 17 18 13 16 14 2 5tv 081308 03...
Page 83: ...Page 34 14 WALL CONTROL ASSEMBLY 57 010307 04 1 5 6 9 10 3 2 7 8 4...
Page 85: ...Page 36 15 WALL CONTROL COMPONENTS 57 010307 05 1 2 3 4 5 6 12 13 14 7 8 9 10 11...
Page 91: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 92: ...MAINTENANCE MANUAL LED SERIES SURGICAL LIGHT 7 07...
Page 93: ......
Page 95: ...Page 2...
Page 110: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 111: ...REV 8 08 INSTALLATION INSTRUCTIONS LFS SERIES SURGICAL LIGHTS...
Page 133: ...Page 21...
Page 135: ...Page 23...
Page 138: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 139: ...LFS SERIES OWNERS MANUAL REV 6 08...
Page 140: ......
Page 142: ...Page 2...
Page 160: ...Page 20...
Page 170: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 171: ...INSTALLATION INSTRUCTIONS AUR SERIES SURGICAL LIGHTS TEC B 0006 REV1 8 09...
Page 188: ...Page 16 TYPICAL AURORA AUR LIGHT FIXTURE CONDUIT AND WIRING REQUIREMENTS...
Page 189: ...Page 17 TYPICAL WIRING DIAGRAM...
Page 190: ...Page 18 TYPICAL AURORA AURTV LIGHT FIXTURE CONDUIT AND WIRING REQUIREMENTS...
Page 191: ...Page 19 TYPICAL WIRING DIAGRAM...
Page 196: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...
Page 197: ...AUR SERIES SURGICAL LIGHTS OPERATORS MANUAL TEC B 0003 REV9 3 10...
Page 198: ......
Page 219: ......
Page 220: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 221: ...PARTS CATALOG AUR SERIES SURGICAL LIGHT TEC B 0004 REV6 6 10...
Page 222: ......
Page 274: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 275: ...AUR SERIES SURGICAL LIGHTS MAINTENANCE MANUAL TEC B 0007 REV1 1 10...
Page 276: ......
Page 278: ...Page 2...
Page 292: ...Page 16 TYPICAL WIRING DIAGRAM...
Page 293: ......
Page 294: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 616 656 2900 FAX 616 656 2906...
Page 295: ...TEC B 0008 REV3 3 10 OWNERS MANUAL HANDLE CAMERA SYSTEM...
Page 296: ......
Page 308: ...Page 12 PRECISION HD CAMERA SYSTEM...
Page 310: ...5085 Corporate Exchange Blvd S E Grand Rapids MI 49512 1 616 656 2900 FAX 1 616 656 2906...