6-6. Switch-Over Relay
a. Switch-Over Relay in OFF Position
The Switch-Over Relay supplies the 24 volt input
power from either the BATTERY or AC120V oper-
ating modes to the relay box for table operation. In
the normal OFF position, BATTERY power is sup-
plied to the relay box. See figure 6-9.
Figure 6-9. Relay in OFF Position
b. Switch-Over Relay in Activated Position
When the AC120V mode is activated by the main
switch, a signal from the relay box activates the
Switch-Over Relay. The relay then supplies the
AC operating mode output power to the relay box
and also activates the battery charging circuit. See
figure 6-10.
NOTE
The battery charging circuit is only op-
erational when the table is in the
AC120V operating mode.
6002B
CN15
CHARGING
BOX
CN14
(-)
B)
CAPACITOR
FUSE
15A
RELAY BOX
RELAY BOX
SIGNAL OUT
S-O RELAY
2
1
3
4
8
6
5
7
CN4
Figure 6-10. Relay in Activated Position
c. Switch-Over Relay Test
Using a DC voltmeter, test the operation of the
relay in both the OFF (AC120V- OFF ) and Acti-
vated (AC120V - ON ) positions. See figure 6-11.
NOTE
The Switch-Over Relay mounting block
may have to be removed from the base
for test access.
OFF: (AC120V - OFF)
term. 7(-) and term. 1(+) = 24 to 28VDC
term. 7(-) and term. 6(+) = 0VDC
Activated: (AC120V - ON)
term. 7(-) and term. 6(+) = 26.5±1VDC
2
7
1
8
3
4
5
DCV
6
Figure 6-11. Switch-Over Relay
Summary of Contents for ELITE 6002
Page 2: ...Page 52...
Page 8: ...Page 52...
Page 29: ...Page 21 Figure 4 1 6002 Electrical Circuit Block Diagram...
Page 59: ...Page 51 VR2 VR3 VR1 Figure 7 2 Relay Box Adjustments Model 6002B...
Page 60: ...Original printing 5 03 Revised 8 05...
Page 61: ...Page 52...
Page 62: ...Page 58 500036thStreetS E GrandRapids MI 49512 1 800 SKYTRONor1 616 957 0500 FAX1 616 957 5053...