INTRODUCTION

INTRODUCTION

This product uses the Oscillometric Measuring Method to detect blood pressure.

Before every measurement, the unit establishes a “zero point” equivalent to the

atmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit detects

pressure oscillation generated by beat-to-beat pulsatile, which is used to 

determine the systolic pressure and diastolic pressure as well as pulse rate.  

Measurement Principle

1. The device should not  be used by any person who is or may be pregnant. 

2. The device is not suitable for use on patients with implanted electrical devices, 

such as cardiac pacemakers and defibrillators. 

The signs below might be in the user manual, labeling or other components. 

They are the requirement of standard and using. 

Safety Information

Symbol for “RECYCLE”

Symbol for “THE OPERATION 

GUIDE MUST BE READ”

Symbol for “MANUFACTURER”

Symbol for “SERIAL NUMBER”

Symbol for “TYPE BF APPLIED 

PARTS”

Symbol for “DIRECT CURRENT”

Symbol for “ENVIRONMENT 

PROTECTION - Electrical waste 

products should not be disposed of 

with household waste. Please recycle 

where facilities exist. Check with your 

local authority or retailer for recycling 

advice”

Caution: These notes must 

be observed to prevent any 

damage to the device.

SN

Symbol for “MANUFACTURE

DATE”

General Description

The iBloodPressure

®

 monitor features blood pressure measurement,

 

pulse rate measurement, and cellular result transmission to cloud

 

storage. The design provides you with two years of reliable service.

 

Readings taken are equivalent to those obtained by a trained observer

 

using the cuff and stethoscope auscultation method. This manual 
contains important safety and care information and provides step by 
step instructions for using the product. Read the manual thoroughly 
before using the product.

Features:

78*92 mm Digital LCD display 

Measuring during inflation technology

E-MTC wireless communication

Indications for Use

The iBloodPressure Blood Pressure Monitor is digital monitors intended for 

use in measuring blood pressure and heartbeat rate with arm circumference 

ranging from 22 cm to 42 cm (about 8.6 in - 16.5 in) or 22 cm to 45 cm 

(about 8.6 in. – 17.7 in.).

It is intended for adult indoor use only.

Contraindications

Warranty

Smart Meter LLC warrants that this product will be free from defects in materials 

and workmanship for two years from the date of purchase.

This warranty does not apply to the performance of an iBloodPressure

®

Meter that has been altered, misused, tampered with or abused in any way. This 

warranty applies only to the original purchaser of the iBloodPressure. 

iBloodPressure is a trademark of Smart Meter LLC

Summary of Contents for iBloodPressure SMBP802-GS-001

Page 1: ...Model SMBP802 GS 001 User Manual To use the monitor correctly and safely please read the manual thoroughly Please keep this manual in order to reference in future Manufactured for Smart Meter LLC 201...

Page 2: ...a Measurement INFORMATION FOR USER 14 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 16 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Ir...

Page 3: ...revent any damage to the device SN Symbol for MANUFACTURE DATE General Description The iBloodPressure monitor features blood pressure measurement pulse rate measurement and cellular result transmissio...

Page 4: ...presence of a flammable anesthetic mixture with air of oxygen or nitrous oxide This unit is not suitable for continuous monitoring during medical emergencies or operations INTRODUCTION INTRODUCTION T...

Page 5: ...onments may affect the performance of the unit Please use a soft cloth to clean the whole unit before and after use Don t use abrasive or volatile cleaners Do not wash the cuff in a washing machine or...

Page 6: ...ims The display does not light up Do not use new and used batteries together Do not use different types of batteries together Do not dispose the batteries in fire Batteries may explode or leak Remove...

Page 7: ...so that the tube exits towards the hand Place your arm through the cuff loop with your palm facing up Position the cuff s edge about an inch 2 3 cm above the elbow Align the marker located to the rig...

Page 8: ...links on the LCD indicates data is transmitting Taking a Measurement Transmitting results Tip You can press START STOP button at any time to stop measuring during the process 3 If the data is successf...

Page 9: ...y place and avoid the sunshine Avoid intense shaking and dropping Use a damp cloth to remove dirt Avoid contact with water clean exterior with a dry cloth Avoid dusty places and unstable temperatures...

Page 10: ...t the day due to weather emotion exercise etc Also there is the white coat effect which means blood pressure usually increases in clinical settings It is okay to measure on either arm however there ma...

Page 11: ...is allowed Power supply Display mode Measurement mode Oscillographic testing mode Measurement range Measurement perimeter of the upper arm Weight External dimensions Attachment Mode of operation Cont...

Page 12: ...device labels labelling and information to be supplied Part 1 General requirements IEC 80601 2 30 2018 Medical electrical equipment Part 2 30 Particular requirements for the basic safety and essential...

Page 13: ...d amateur radio bands between 0 15 MHz and 80 MHz 6V For AC power port 3 Vrms 6 Vrms in ISM and amateur radio bands 80 AM at 1 KHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1...

Page 14: ...Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if...

Reviews: