EMC GUIDANCE

SPECIFICATIONS

EMC Guidance

The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments

Warning: Do not use near active HF surgical equipment and the RF shielded room of an ME system for 
magnetic resonance imaging, where the intensity of EM disturbances is high.

Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it 
could result in improper operation.  If such use is necessary, this equipment and the other equipment should be 
observed to verify that they are operating normally.

Warning: Use of accessories, transducers and cables other than those specified or provided by the 
manufacturer of this equipment could result in increased electromagnetic emissions or decreased 
electromagnetic immunity of this equipment and result in improper operation.

Warning: Portable RF communications equipment (including peripherals such as antenna cables and external 
external antennas

)

 should be used no closer than 30 cm (12 

inches)

 to any part of the equipment SMBP802-

GS-003 or SMBP802-GS-004, including cables. Otherwise, degradation of the performance of this equipment
could result.

Technical description

1

All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard  

to electromagnetic disturbances for the excepted service life.

2

Guidance and manufacturer’s declaration -electromagnetic emissions and immunity

Table 1 

Guidance and manufacturer’s declaration - electromagnetic emissions

RF emissions 

CISPR 11

Group 1 

Class [ B ]

Class A

Comply

Compliance

Harmonic emissions 

IEC 61000-3-2

Voltage fluctuations/ 

flicker emissions  

IEC 61000-3-3  

RF emissions 

CISPR 11

Emissions test

 

IP21: It means the device could be protected 

against solid foreign objects of 12,5mm Ф and 

greater, and against vertically falling water drops.

Protection against 
water ingress 

Device Classification

Battery Powered Mode: 
Internally Powered ME Equipment 
AC Adapter Powered Mode: 
Class II ME Equipment

General Environment

- home use, not intended for professional use
- indoor use only
- not for use in shower, bath tub or sink

Normal working 
condition

A temperature range of: +5°C to +40°C 
A relative humidity range of 15% to 90%,
non-condensing, but not requiring a water vapour
partial pressure greater than 50 hPa
An atmospheric pressure range of: 
700 hPa to 1060 hPa     

Storage & transportation 
condition

Temperature: -20°C to +60°C

A relative humidity range of ≤ 93%, 

non-condensing, at a water vapour pressure 
up to 50 hPa 

Conditions of visibility

Ambient luminance range: 100 Ix to 1500 Ix

≤50 dB in the range of 100 Hz — 8 kHz

Background sound 
pressure level

Software Version

A01

Service life

2 years

WARNING: No modification of this equipment is allowed.

Summary of Contents for iBloodPressure SMBP802-GS-003

Page 1: ...04 User Manual To use the monitor correctly and safely please read the manual thoroughly Please keep this manual in order to reference in future Manufactured for Smart Meter LLC 5501 W WATERS AVE STE...

Page 2: ...a Measurement INFORMATION FOR USER 14 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 16 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irr...

Page 3: ...rate with arm circumference ranging from 22cm to 45cm about 8 17 The device can be used to detect irregular heartbeat It is intended for adult indoor use only Contraindications Warranty Smart Meter L...

Page 4: ...al emergencies or operations INTRODUCTION INTRODUCTION To avoid measurement errors avoid electromagnetic field radiated interference signals or electrical fast transient burst signals Wireless communi...

Page 5: ...vironments may affect the performance of the unit Please use a soft cloth to clean the whole unit before and after use Don t use abrasive or volatile cleaners Do not wash the cuff in a washing machine...

Page 6: ...lay dims The display does not light up Do not use new and used batteries together Do not use different types of batteries together Do not dispose of the batteries in fire Batteries may explode or leak...

Page 7: ...so that the tube exits towards the hand Place your arm through the cuff loop with your palm facing up Position the cuff s edge about an inch 2 3 cm above the elbow Align the marker located to the rig...

Page 8: ...on the LCD indicates data is transmitting Taking a Measurement Transmitting results Tip You can press START STOP button at any time to stop measuring during the process 3 If the data is successfully t...

Page 9: ...y place and avoid the sunshine Avoid intense shaking and dropping Use a damp cloth to remove dirt Avoid contact with water clean exterior with a dry cloth Avoid dusty places and unstable temperatures...

Page 10: ...t the day due to weather emotion exercise etc Also there is the white coat effect which means blood pressure usually increases in clinical settings It is okay to measure on either arm however there ma...

Page 11: ...fasten the cuff and then measure again The measurement process does not detect the pulse signal Loosen the clothing on the arm and then measure again Not able to calculate measurement failed Relax for...

Page 12: ...RMANCE with regard to electromagnetic disturbances for the excepted service life 2 Guidance and manufacturer s declaration electromagnetic emissions and immunity Table 1 Guidance and manufacturer s de...

Page 13: ...ds between 0 15 MHz and 80 MHz 6V For AC power port 3 Vrms 6 Vrms in ISM and amateur radio bands 80 AM at 1 KHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Radiated RF IE...

Page 14: ...interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following mea...