background image

 

 

PowerForce 

 User Manual 

Radlabor GmbH | Heinrich-von-Stephan-Str. 5c | 79100 Freiburg | Germany 
www.smartfit.bike | www.radlabor.de 

 

1.2 Force components 

The PowerForce records the forces, which are performed on the pedal axis, separately for each leg. 
Three force components (Fe, Fu und Fr) have to be differentiated. 
 

 

Fr (resultant / total 

 force) 

This is the total force which is generated by the athlete on the pedal. It is the vectorially sum of 

Fe

 

and 

Fu

 an is calculated, not measured. 

Fr

 tells us about the total load, the rider applies to the system 

 the bike as well as to his own body. 

Joint moments and forces result from (or cause) this component. Thus, e.g. inverse dynamics are 
based on 

Fr

 

Fe (effective / tangential 

 force) 

Fe

  is  the  component  of 

Fr

  which  lies  in  moving  direction  of  the  crank/pedal  which  is  a  circle. 

Therefore, the vector of 

Fe

 is always tangential to this circle or perpendicular to the crank.  

Fe

 is 

responsible  for  the  propulsion,  but  only  if  it  is  aligned  forward  with  the  rotating  direction.  If  it  is 
aligned backward, e.g. if a rider does not pull up the pedal in the upstroke phase, 

Fe

 of the one side 

must be subtracted from the other side. 
 

Fu (unused / radial 

 force) 

Fu

  is  the  component  of 

Fr

  which  does  not  generate  any  propulsion.  It  results  in  very  small 

deformation of the crank and pedal-axis. The direction off this force component is lengthwise to the 
crank.  The  highest  values  appear  around  the  bottom  dead  center  (180°  crank  angle)  where  the 
weight component (induces by the mass of the leg) adds to the muscular component. 
 
 
 
 
 
 
 

Summary of Contents for PowerForce

Page 1: ...PowerForce User Manual Radlabor GmbH Heinrich von Stephan Str 5c 79100 Freiburg Germany www smartfit bike www radlabor de PowerForce System User manual wireless edition ...

Page 2: ... unpacking 8 2 1 components 8 2 2 Mounting 9 3 Software Installation 12 3 1 Install NI DAQmx Driver 12 3 2 Install NI LabVIEW 2009 Runtime Engine Library 12 3 3 Install IMAGO Records 12 4 IMAGO Record 13 4 1 Introduction 13 4 2 First start and setup 13 4 2 1 Load Setup files 14 4 2 2 Set save and load calibration values 15 4 3 Description of IMAGO 17 4 3 1 Main Menu 17 4 3 2 Index Tab 18 4 4 Measu...

Page 3: ...itter unit or the connection box the warranty is no longer valid Never clean the sensor or the transmitter unit with water or use it in wet conditions Do not cramp the position switch it is sensitive for pressure Check before each ride that the sensors and the pedals are fixed properly When unplugging any connection always pull straight down rather than sideways or you could damage connector pins ...

Page 4: ...rerequisite to analyze cycling performance capability from a biomechanical point of view The question is not why to measure pedal forces but how to do it Comparing other existing instruments whose purpose is the determination of divers mechanical parameters concerning cycling propulsion there are some methodological or practical limitations regarding the requirements of either a biomechanical lab ...

Page 5: ...e based on Fr Fe effective tangential force Fe is the component of Fr which lies in moving direction of the crank pedal which is a circle Therefore the vector of Fe is always tangential to this circle or perpendicular to the crank Fe is responsible for the propulsion but only if it is aligned forward with the rotating direction If it is aligned backward e g if a rider does not pull up the pedal in...

Page 6: ...nk The system determines the magnetic field variations Hall Effect because of the angular displacement 1 1000mm of a small magnet in respect to a sensor Due to the specific circular construction of the force transmitting elements the applied force can be divided into a radial and a tangential part Positive tangential force FT Fe is in the pedaling direction and positive radial force FR Fu is in th...

Page 7: ... Data Wireless edition Mass of single sensor including radio 370 g Mass of receiver 116 g Measurement range 500 up to 1500 N Breaking load 3000 N Sampling frequency pedal force 500 Hz Wireless band ISM Band 2 4 GHz Transmission distance Indoor 3 5 m Outdoor 2 3 m Battery Type NiCd 800 mAh Time for continual use 6 8 h Resolution 12 Bit ...

Page 8: ...rts with highest care 2 1 components Check the content inside the suitcase 1 2x PowerForce sensor left right with mounted transmitter Accumulator included two protector caps mounted 2 1x Base station receiver 3 1x Adjustment Tool 4 2x adjustment tools screws 5 2x crank sensor screws 6 2x Accumulator Charger 7 1x Serial USB cable Base station to PC 8 1x USB License Dongle 9 1x 10mm Allen Key 10 1x ...

Page 9: ...w smartfit bike www radlabor de 2 2 Mounting To improve the measurement it is important to fix the sensor in appropriate position on the ankle Before the mounting make sure that the adjustment screws are screwed out so that the sensor has some space to move Then mount the sensor on the crank ...

Page 10: ...many www smartfit bike www radlabor de Adjust the apex on the cone bolt Put the little bolts in the middle and fix the adjustment tool hand tight onto sensor by using the corresponding screw for each site Put the cone of the adjustment tool into the center of the bottom bracket screw ...

Page 11: ...ustment screws Tighten the screws alternating It means first a little the left then a little the right until the screws fixing the sensor in position Make sure that the cone of the adjustment tool stays in the center of the bottom bracket screw After the adjustment screws fixing the sensor tight the sensor crank screw Then you can take off the adjustment tool ...

Page 12: ...nstall the NI DAQmx Driver Click on the icon in the saved file and follow the setup instructions The Installation will take some minutes 3 2 Install NI LabVIEW 2009 Runtime Engine Library After installing the DAQmx Driver you must install the NI LabVIEW Runtime Engine This library is necessary to read out the data from IMAGO Click on the icon and follow the setup instructions The Installation will...

Page 13: ... PowerTec pedal force measurement system and EMG recordings For training and training studies the program provides an online feedback This feature called Cycle Plot simplifies the data stream in a circular graph to enable any athlete to work with feedback without previous knowledge As the program has many more features used at the University which are not available in your version there might be a...

Page 14: ...main record window IMAGO record vi and a window with the overview of all channels Wireless Cycling Instrument vi 4 2 1 Load Setup files To load the optimum setup files you must click on setup Find the folder record on the following data path Then you can insert the downloaded setup file called RECORD_CYCLE1 set After this step the optimal set up settings should be present in IMAGO ...

Page 15: ...u can see the offset settings and the correction values that were determined during the calibration of your system in the Radlabor Company Now enter the for your system individual parameters in the Wireless Cycling Instrument VI as seen in the picture Note that the gain for the power channels must be calculated with 500 𝑁 𝑉 to get from the voltage in V to the force in N Then press on the save butt...

Page 16: ...t bike www radlabor de To have these settings at each start of IMAGO you need these set files loaded in the Read Write Global Variables VI You find this VI under the path Options Extras Paths Then you must load the set files as shown in the picture After these adjustments your system is ready to use ...

Page 17: ...y To stop this automatism you need to switch the Always Active option Further here is the setting the page and for printing the actual window Beenden means to close IMAGO Tools Via the Tab Tools you can do the Binary Export You need this binary export to transform your binary data in an excel readable file format e g for analyzing the separated force channels or the acceleration channels Extras He...

Page 18: ...elp Via the Help tab you have three selection possibilities If you select WOZU KONTEXTHILFE If you select About IMAGO you can check which version of IMAGO you have Further you can see the key number of your USB license dongle Under User Manual you can call up further usage instructions for IMAGO Records To this your PC must be connected to the internet 4 3 2 Index Tab Via the Index Tab you can cho...

Page 19: ... start angle the crank length and the used data pieces Index tab Wireless Before you start the measurement it is recommended to check the wireless connection of your PowerForce system So you have an overview about the current wireless signals in V the base station receives Under the wireless signals you see an overview about the arriving frequency samples Usually they are about 500 Hz 1 When you j...

Page 20: ...H Heinrich von Stephan Str 5c 79100 Freiburg Germany www smartfit bike www radlabor de If you notice a high increasing number of interrupts it may occur because of an interference of your Wi Fi Signal or a damage on the transmitter unit ...

Page 21: ...the bigger is the measured negative force The displayed channels can be changed above the charts The power calculations depend on the crank length which has to be entered below the cycle plot window If the Trigger is not installed at the top dead center of the right crank arm the offset can be entered as well To start data recording click the RECORD button You can pause data recording by clicking ...

Page 22: ...007 but an older Excel 2003 Version is also available which is containing less functions The Tool is written in Visual Basic for Applications When you open CyclePlotAssistant xltm the first step you need to take is to activate Macros for this Workbook You start the CyclePlotAssistent by click on the Icon Make sure that the macro files are in the same folder as the application So its recommended to...

Page 23: ...f CPA you see different icons If you click on Main_Menu you can open the menu By clicking on the arrows above Zero circle you can change the size of the zero circle for cycle and vector plots It Has no relevance for other charts In the box of graph size you can change between maximal amplitude for all graphs or the same proportion standard ...

Page 24: ... buttons and contents The file icon opens the xls files you want to display they must be recorded with IMAGO The white cross on red background next to the file icon unload the selected files The blue numbers are quick load buttons Once you have created a sheet you can save your settings and apply it to other data with these buttons ...

Page 25: ...is divided into six areas where charts can be plotted There are also description fields available If you click on Corner Caption or Header you can arrange the header and the headline of the print sheet The information is taken from the loaded raw data files Multiple choices are impossible Cycle In this chart the data is plotted in a circle The black circle is zero Positive data is outside the blac...

Page 26: ...position Note that the first coordinate is always the direction of Fe the second always that of Fu Two different channel pairs can be shown in one graph XY Graph 5 3 Graph Selection Here you can see which file is represented by data1 data2 data3 and data4 The button details open the detail window This appears in the right part of this window and all selected graphs are shown there To create new Qu...

Page 27: ...one chart 5 4 Caption and legend Add manually Comments which can be added to the header label or legend 1 Similar to the overview for the graph selection an overview window for the top bar and legend labels exists The left list represents the Top bar selection the right one the legend selection Multiple choices are possible 2 By clicking on the chart position you can apply a name to the Chart in t...

Page 28: ...on all Indices of the current selection charts comments graphs legends and header entries are saved Note that only the indices are saved Meaning that if you have selected the second recorded channel of your file the second channel of the new selected data you apply your settings to will be shown There is no reference to the selected channel name By clicking on New Button Name you can change the na...

Reviews: