80
Limitations of the measurement procedure
1. DO NOT use serum or plasma sample.
2. DO NOT use neonate blood sample.
3. DO NOT use anticoagulant NaF or potassium oxalate for venous
sample preparation.
4. Extreme humidity may affect the results. A relative humidity
greater than 90% may cause incorrect results.
5. The system is designed to be used at temperatures between 10°C
and 40°C (50°F and 104°F). Outside this range, the system may
yield erroneous results.
6. DO NOT reuse the test strips. The test strips are intended for
single use only.
7. DO NOT use iodoacetic acid, fluoride or sodium fluoride/oxalate
as a preservative for blood specimens.
8. Hematocrit:
Test strip results are not significantly affected by hematocrits in
range of 30% to 55%. Hematocrit levels less than 30% may cause
incorrect high readings and hematocrit levels greater than 55%
may cause incorrect low readings. If you do not know your hema-
tocrit level, consult your healthcare professional.
9. Critically ill patients should not be tested with
smartLAB®
nG
Blood Glucose Systems .
10. DO NOT use during xylose absorption testing. Xylose in the blood
will interfere Self-Monitoring Blood Glucose System.