RotaScope Classic Cabinet
Manual Version
1.04
Date
23 November 2021
85
6
MAINTENANCE
The RotaScope shall only be maintained by a qualified and trained Authorised
Distributor
How to Maintain the Product
6.1.1
Planned maintenance of RotaScope.
Maintenance tasks shall be done according to the following plan:
6.2
How to Inspect the Product
6.2.1
Weekly inspection tasks
Task
Frequency
Replacing the Intake Filter
Every 12 months
Replacing HEPA Filter
Every 12 months
Replacing the Air Pumps
Every 24 months
Task
Action
Weekly Cabinet Clean
Use Hospital Grade Disinfectant to
wipe over every internal surface of the
cabinet as per the IFU in this manual
Weekly Scope Hose Connection Set
Clean (Clean Hoses or swap-out with
a previously cleaned hose kit)
Remove and replace hoses from
cabinet and have them disinfected
using the manual or automated
disinfection process as per the IFU in
this manual