MAN-0901 RotaScope
VALID DATE: 01/2019 to 12/2020
Model: RotaScope G & Upgrades
18
© Copyright Smartline Medical Pty Ltd All rights reserved
SPECIFICATIONS TABLE
Topic
Answer
Notes
List of endoscopes that can be stored in
the storage cabinet;
See: Approved Scope List.
Endoscope accessories (e.g. valves) can be
stored inside the storage cabinet but it is not
the intention of the storage cabinet to
maintain the microbiological quality of these
accessories.
Storage temperature band
10
–
40degC
Maximum storage time
168hrs
7 days maximum storage
Drying function
N/A
Cabinet is not equipped with a drying function
Drying temperature band
N/A
Not Applicable as cabinet does not use
elevated temperatures to dry load.
Time required to dry the endoscopes (if
applicable)
3 Hours (MAX)
Dryness may be achieved in less than 3 hours
Channel aeration
Tested at IQ, OQ and PQ Intervals (Install
and Annual PQ)
Testing with use of Smartline Medical Aeration
Testing Device inline between scope channel
and vacuum manifold to check air circulation
through different channels.
Manifold Test with Oxyview
Automated channel flushing control
N/A
Automated channel flushing NOT used
Air Flow circulation diagram
See: USER Manuals
See: Benefits of Shared Air System
RotaScope:Page 10
SlidaScope: Page 10
Cabinet Flow Rate (Ports at Manifold)
>1.0 l/m
Minimum detectable flow rate is .5 l/m
Cabinet Flow Rate (Total)
200 l/m
10x Air Exchanges p/h
Minimum detectable flow rate is 20 l/m
Total Vacuum Flow Rate Per Cabinet
170l/m
Cabinet Pressure Band
-1
–
27pa
Differential Pressure (above 27Pa is possible
but not recorded.
Endoscopes that one or more channels
is below the detection limit
N/A
Currently there are no endoscopes that can be
stored in the cabinet that have a flow rate that
is below the detectable flow rate of 1 l/min.
Air Quality
99.99% @0.3µm Filtration Rate (HEPA
Filtered Supply Via SAB)
Third party air is not used by the storage
cabinet. A Shared Air Box draws required air
from the surrounding environment of which it
is situated and therefore standard office
working conditions need to be provided at all
times. Room temperature range 10 C° to 30 C°,
relative humidity between 40% and 70%,
environmental room pressure greater than 102
kPa.
Humidity Band
3.0
–
90.0 %RH
Oil Content
N/A
Compressed air not used
Air Quality Testing Frequency
Annually
–
Service and PQ
HEPA Filter replaced Annually