Peltier-type Chiller

Thermo-con/Compact Type 

Air-cooled

HEF

 Series

210

mm

of a heat source or process fluid

Construction and Principles

Heat dissipation air fl ow direction

The thermo-con is constructed as shown in Figure 1 . The circulating fl uid outlet temperature 

is controlled precisely by a Peltier element (thermo module) between the heat exchanger 

and heat sink which is controlled by supplying DC power supply. The circulating fl uid returns 

to the tank and is transferred by the pump which is built into the thermo-con, then it goes 

through the heat exchangers and temperature sensor and out from the circulating fl uid outlet.

The heat dissipation air fl ow direction is as 

shown in Figure 2. Air is taken from the front 

of the Thermo-con and discharged from the 

rear. Do not cover the inlet and outlet vents.

Integrated with the front panel 
grille. Easy to install and remove

Easy to remove dust, cutting 
chips, etc., stuck to the fi lter with 
a brush or vacuum cleaner

Dustproof fi lter set

Dustproof fi lter

p. 

11

Easy maintenance

Easy to clean and replace the 
dustproof fi lter

Air

Air

p. 

11

T

2

Peltier device T 
(Thermo-module)

Controller

Figure 1

Figure 2

Tank

Pump

Heat sink

Heat exchanger

Fan

Level switch

Power switch

Power supply 
(24 VDC)

Circulating fl uid
OUT

Temperature 

sensor

Circulating fl uid 
fi ll port (Tank lid)

Circulating fl uid
IN (Drain)

Summary of Contents for HEF Series

Page 1: ...94 s a 41 s Compared with the existing product HECR002 Time to change the circulating fluid temperature from 30 C to 20 C when the circulating fluid IN and OUT are directly connected Now only takes ab...

Page 2: ...se design Noise Level Sound level Noise level reference Sound example 20 dB Whisper The sound of leaves touching each other or the sound of snow falling 30 dB Very quiet Suburbs late at night or the s...

Page 3: ...emperature sensor and out from the circulating fluid outlet The heat dissipation air flow direction is as shown in Figure 2 Air is taken from the front of the Thermo con and discharged from the rear D...

Page 4: ...m Bumper Circulation Between Gas Generation Analyzer and JacketedVessel Application Examples Peltier type Chiller Thermo con Compact Type Air cooled HEF Series C mount lens F mount lens Microscope C m...

Page 5: ...ecifications p 6 Cooling Capacity p 7 Heating Capacity p 7 Pump Capacity Thermo con Outlet p 7 Dimensions p 8 Operation Display Panel p 9 Alarm p 9 Maintenance p 9 Option High Pressure Pump Mounted p...

Page 6: ...down to 20 C Water bath Example When cooling the object below a certain temperature in certain period of time Cooled substance total volume V 2 L Cooling time h 15 min Cooling temperature difference...

Page 7: ...4 Fluid contact material Stainless steel EPDM NBR Ceramics PPE PPS Carbon PP POM Electrical system Power supply 24 VDC 10 Power consumption 12 5 A Peak current 18 A Alarm Refer to Alarm Communication...

Page 8: ...rox 20 W For the high pressure pump Option the heating capacity increases by approx 10 W Pump Capacity Thermo con Outlet Circulating fluid Water 0 18 0 16 0 14 0 12 0 10 0 08 0 06 0 04 0 02 0 0 1 1 5...

Page 9: ...ignal input 5 SG 6 Output cutoff alarm open when alarm occurs 7 Common for output cutoff alarm 8 RUN STOP signal input 9 RS 485 BUS Power Cable Accessory Connector J42FSC 02V KX J S T Mfg Co Ltd made...

Page 10: ...be repaired at one of SMC s sites As a rule SMC will not conduct on site maintenance Alarm Alarm code Alarm description Operation status Main reason AL0 Memory error Stop There is an abnormality with...

Page 11: ...customer s piping resistance Cooling capacity is reduced by approx 20 W due to the heat generation from the pump P HEF002 A6 P High pressure pump mounted Options have to be selected when ordering the...

Page 12: ...ecifications Power supply input 100 240 VAC Single phase 50 60 Hz Max 3 9 A DC output 24 VDC Max 12 5 A Operating environment Ambient temperature 0 to 40 C Humidity 20 to 80 RH no condensation DC outp...

Page 13: ...s less than these values the thermo con will not be able to provide precise control and the repeated cooling and heating operations may cause it to fail 7 A magnet driven pump is used as the circulati...

Page 14: ...Copper Cu mg L 0 1 or less Sulfide ion S2 mg L Should not be detected Ammonium ion NH4 mg L 0 1 or less Residual chlorine Cl mg L 0 3 or less Free carbon CO2 mg L 4 0 or less 11 If there is a point w...

Page 15: ...of SMC products with production equipment for the manufacture of weapons of mass destruction WMD or any other weapon is strictly prohibited 2 The exports of SMC products or technology from one countr...

Page 16: ...c hu Ireland 353 0 14039000 www smcautomation ie sales smcautomation ie Italy 39 03990691 www smcitalia it mailbox smcitalia it Latvia 371 67817700 www smc lv info smc lv Lithuania 370 5 2308118 www s...

Reviews: