HRW***-* -Y
Material for DI control kit (optional)
1
2
Circulating fluid return
(Rc3/4)
Circulating fluid supply
(Rc3/4)
DI circuit return
(3/8 self-align fitting/with plug)
DI circuit supply
(3/8 self-align fitting/with plug)
Applicable model
Available circulating fluid for DI circuit
DI level indication range
DI level setting range
Control solenoid valve hysteresis
-
M
Ω・
cm
M
Ω・
cm
M
Ω・
cm
-
0
~
20*
2
0
~
20*
1
HRW***-H1-Y/HRW***-H1S-Y HRW***-H2-Y/HRW***-H2S-Y
60% Ethylene glycol solvent
DI water
Thermo Chiller DI circuit supply
:
Option
Solenoid valve
DI sensor
Circulating fluid tank in Thermo Chiller
Piping for customer's system
Circurating
fluid supply
(Rc3/4)
Circulating fluid
return
(Rc3/4)
Strainer
Ball valve
(IN/OUT)
DI filter
(not attached)
DI circuit
Thermo Chiller DI circuit Return
*This option does not have DI filter.
For details, please refer to specification table and notes.
Facility water
Note
1.This product is designed as an option for Thermo chiller, Series HRW. Do not use this product for the other purpose, or outsi de
of the specification range.
2.Installation of accessories has to be performed by a customer according to the operation manual.
3.Supply circulating fluid from IN side of the DI filter to exhaust air. Continue to supply unitl OUT side is filled with circulati ng
fluid.
4.The DI filter has water inside. If ethylene glycol solvent is used, the concentration may be decreased.
Confirm the concentration after the trial operation. The proper concentration of ethylene glycol is 60 %.
Dimensional tolerance for each: 10mm
Accessories
Specification Table
DI tube for IN/OUT
(with a insulator) x 1pc for ea ch
Operation manual (this documen t)
[document no:HRX-OM-K018]Jpn x 1 copy, Eng x 1 copy
0
~
0.9
*1: DI filter (Part no.: HRZ-DF001) is necessary to control DI level. This optional product does not have DI filter, so
purchase it separately. If this product is used out of the range of 20 to 40degC, also purchase a thermal insulat or
for DI filter (our part no.: HRX-DF002) to avoid frostbite and bur n.
*2: Temperature correction is not applied to the DI level indication valu e.
Note: Temperature stability at supply of circulating fluid in the Thermo Chiller could be out of +/ -0.3degC depending
on DI filter replacement and operating condition s.
Flowchart
DI filter/ Thermal insulator for DI filter
(This option does not have DI filter and Thermal insulator.)
DI filter
Part No.
:
HRZ-DF001
Thermal insulator for DI filt er
Part No.:HRZ-DF002
DI filter/ insulator for DI filter
mounted condition
(380
)
(860
)
(665)
(242)
(294
)
DI level lowering alarm setting range
0
~
20
(120)
(1
4
3
)
M
Ω・
cm
1
2
Applicable model
HRW***-* -Y
HRW***-* S-Y
1
2