The HOTLINE
®
employs a safe recirculating solution heating
system, inherently free of hotspots, to actively warm the HOTLINE
®
Blood and Fluid Warming Set
- a co-axial arrangement of warm
fluid-filled jacket and sterile IV line. The primary temperature
control circuit limits the recirculating solution to 42
o
C maximum. In
the unlikely event of a malfunction of this circuit, a second
watchdog circuit will visually and audibly alarm and stop pumping
the recirculating solution if the temperature reaches 43
o
C.
The heater current is supplied via a thermal cutout, so in the
case of a circulation failure, the heater will be protected against
burn-out; the same cutout will act in the event of catastrophic
breakdown of components in the HOTLINE® circuitry that prevents
the alarm state from disengaging the heater.
These three levels of safety ensure that the IV Fluid in the
HOTLINE
®
Fluid Warming Set is never exposed to any damaging or
dangerous temperature while the unit is operating.
Mechanical
The rear left-hand side of the unit is a reservoir, holding 1.4
litres of either distilled water or a disinfectant solution. This fluid
will henceforth be referred to as recirculating solution. Do not
confuse recirculating solution with the IV Fluid being dispensed to
the patient.
Summary of Contents for level 1 HOTLINE HL-90
Page 1: ......
Page 14: ...Introduction Specifications Blood and Fluid Warmer ...
Page 20: ...Sub Assemblies Blood and Fluid Warmer ...
Page 28: ...Technical Description Blood and Fluid Warmer ...
Page 45: ...Disassembly Procedures Blood and Fluid Warmer ...
Page 58: ...Blood and Fluid Warmer Maintenance Testing and Calibration ...
Page 83: ...Troubleshooting Blood and Fluid Warmer ...
Page 88: ...Spare Parts Blood and Fluid Warmer ...
Page 96: ...Rounding Off Blood and Fluid Warmer ...
Page 106: ...Appendix 1 Circuits Blood and Fluid Warmer ...
Page 110: ...Appendix 2 PCB Layout Blood and Fluid Warmer ...
Page 115: ...Appendix 3 Symbols Glossary Blood and Fluid Warmer ...
Page 118: ...Appendix 4 Frequently Asked Questions Information Bulletins Blood and Fluid Warmer ...
Page 132: ...Ndjg CdiZh 9 V gVbh ...