S E C T I O N 7
•
C o n t r o l s a n d D i s p l a y s
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
17
Fluid Warmer Display Panel
The Fluid Warmer Display Panel provides continuous information
about the operation of the Fluid Warmer. A liquid crystal display (LCD)
indicates recirculating solution temperature. Just below the LCD, four
light-emitting diodes (LEDs) indicate operation modes for the device.
For identification purposes, the diodes are shown illuminated.
1 Recirculating Solution Temperature
The temperature of the recirculating solution is displayed in the
LCD panel. The temperature is displayed in degrees Celsius.
Note: This is NOT the temperature of fluid delivered to the
patient—the display reflects the temperature of the recirculating
solution.
2 Automatic Operation LED
This green LED indicator illuminates when the power is ON and
the Disposable Set has been properly installed. When lit, this
indicates the Fluid Warmer is operating.
3 Check Disposables LED
This yellow LED indicator illuminates and an audible attention
signal activates when the Disposable Set is not properly installed.
See “Interlocks” on page 19 for directions on clearing the Check
Disposables alarm.
4 Add Recirculating Solution LED
This yellow LED indicator illuminates and an audible attention
signal activates when reservoir is low. Additional recirculating
solution must be added to the reservoir. Maximum capacity for
the reservoir is 1.4 liters.
5 Over Temperature LED
This red LED indicator illuminates and an audible warning signal
activates when the recirculating solution is over the acceptable
temperature for safe use.