S E C T I O N 1 2
•
M a i n t e n a n c e
44
Level
1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
Change O-Rings
Change the O-Rings in the #1 Block and #2 Block.
1
Remove each O-Ring from its socket by pulling it out with a pair
of needle-nose pliers or by prying it out with a small screwdriver.
2
Lubricate the new O-Rings from the O-Ring Kit.
3
Press each O-Ring into its socket.
Clean Fan Filter
The Fan Filter
(a
) is located on the bottom of the Fluid Warmer. The
fan guard snaps in place.
1
Unsnap and remove the fan guard from the bottom of the unit.
2
Remove the filter and clean it with warm soapy water.
3
When the filter is fairly dry, replace the fan guard and filter.
Testing Fluid Warmer Operation
Perform all the tests described in the testing section of this manual. See
“Testing” on page 37.
• Add Recirculating Solution Alarm
• Check Disposables Alarm
• Over Temperature Alarm
• Performance Testing
a