2
HOTLINE
®
Fluid Warmer (HL-90)
HOTLINE
®
Fluid Warming Sets (L-70)
Step-by-Step Guide
•
Plug the Twin-Tube Connector on the HOTLINE
®
fluid
warming set into the socket
•
Connect the IV fluid and the intravenous administration set to
the HOTLINE
®
fluid warming set
•
Fully prime the intravenous administration set and the
HOTLINE
®
fluid warming set prior to connecting to the patient
•
After use, turn OFF the power switch
•
Remove the HOTLINE
®
fluid warming set and insert the reflux
plug into the socket
•
Turn ON the power switch
•
Inspect the patient end of the HOTLINE
®
fluid warming set for
leaks to confirm the integrity of the intravenous pathways
IMPORTANT: Please refer to the Operator’s Manual supplied
with the product for detailed instructions, contraindications,
warnings and cautions and the appropriate IFU for specific
information on using disposable sets.
Warning: Prime the recirculating solution path before
connecting to the intravenous extension set, to confirm there
is no breach between the recirculating solution path and the
intravenous path.
•
1
•
3
•
4
HL-90
L-70