Chapter 12: Service Menu
Veterinary 3 Parameter Advisor
®
Operation Manual
12-3
To restore the factory default alarm limits for each parameter:
See
Using the Static Limits Feature
in this section for details about how turning on or off the static limits feature
affects default alarm limits.
Select the desired patient type (CAT, DOG, or HORSE). The patient type indicator is located on the upper
1.
portion of the display.
To change the patient type:
•
Turn the rotary knob on the monitor to move the cursor. Highlight HORSE, DOG, or CAT, whichever is
a.
displayed. Push the knob to access the patient type submenu in the lower left corner of the display.
Highlight PATIENT TYPE and push the knob to select.
b.
Turn the rotary knob to choose the desired type (HORSE, DOG, or CAT) and push the knob to select.
c.
On the service menu, highlight MONITOR DEFAULTS and push the knob to access the submenu for default
2.
parameter alarm limits.
Highlight DEFAULT PARAMETER ALARM LIMITS and push the knob to select.
3.
Highlight RESTORE FACTORY LIMITS AS UNIT DEFAULTS and push the knob to select.
4.
Highlight YES and push the knob to select.
5.
Highlight MAIN or PREVIOUS and push the knob to select.
6.
Set the ECG Filter Level
You can change the ECG filter level (50 Hz or 60 Hz) depending on the AC line frequency at your facility, so that
each time you turn on the monitor or change the patient type, the ECG filter level will be appropriate for your
location.
To change the ECG filter level:
On the service menu, highlight MONITOR DEFAULTS and push the knob to access the submenu for default
1.
parameter alarm limits.
Highlight ECG FILTER and push the knob to select.
2.
Turn the rotary knob on the monitor to choose the ECG filter level (50 Hz or 60 Hz) appropriate for your
3.
facility and push the knob to select.
Highlight MAIN or PREVIOUS and push the knob to select.
4.