Chapter 1: Introduction
1-4
Veterinary 3 Parameter Advisor
®
Operation Manual
WARNING! Equipment is protected against defibrillator discharge. Rate meters and displays may be
temporarily affected during defibrillation, but will rapidly recover.
WARNING! The vital signs monitor is suitable for use within the patient environment IEC 60950 approved
equipment must be placed outside of the patient environment. The patient environment is
defined as any volume in which intentional or unintentional contact can occur between the
patient and parts of the system or between the patient and other persons touching parts of
the system.
Figure 1.1: Patient Environment (side)
The dimensions are not prescriptive.
Figure 1.2: Patient Environment (top)
The dimensions are not prescriptive.
WARNING! When connecting this monitor to any instrument, verify proper operation before clinical use.
Use only equipment meeting specifications given in this manual. Refer to the instrument’s
user manual for full instructions. Accessory equipment connected to the monitor’s data
interface must be certified according to the respective IED standards, i.e. IEC 60950 for data
processing equipment or IEC 60601-1 for electro medical equipment. All combinations
of equipment must be in compliance with IEC 60601-1-1 systems requirements. Anyone
connecting additional equipment to the signal input port or signal output port configures a
medical system, and therefore, is responsible that the system complies with the requirements
of the system standard IEC 60601-1-1.
WARNING! Any monitor that has been dropped or damaged should be checked by qualified service
personnel to ensure proper operation prior to use.
WARNING! Use only original manufacturer or recommended patient cables. Use of accessories other than
those specified may result in increased electro-magnetic (EM) emissions or decreased EM
immunity of the device. To avoid potential electro-static discharge interference, do not use
cables that incorporate metal or metal-coated connectors.
WARNING! Medical electrical equipment, including this device, needs special precautions regarding
electro-magnetic compatibility (EMC) and needs to be installed and put into service
according to the EMC information provided in this manual.