Chapter 3: Controls and Features
3-4
Veterinary 3 Parameter Advisor
®
Operation Manual
Keys
DESCRIPTION
INSTRUCTION
ON/OFF
Press this key to turn on the monitor. To turn off the monitor, press the key and a message
will be displayed indicating that if you continue to power down, some configuration data
will be lost; press the key again to turn off the monitor.
ZERO IBP
NOT USED
NIBP
Press this key to activate an immediate non-invasive blood pressure (NIBP) measurement.
To cancel an NIBP measurement in progress, press the key again.
NOT USED
ALARM SILENCE Press the key momentarily to silence alarms for 2 minutes. To reactivate the audible alarm
tones within the two minutes, press the key again. Press the key for 3 seconds to silence
alarms indefinitely. AUDIO PAUSED will be displayed. A new alarm will cancel silence. Press
the key for 6 seconds to turn alarm sounds off permanently. AUDIO OFF will be displayed.*
Rotary Knob
The rotary knob is a dial control with a push selection switch. It is located on the front of
the monitor, in the lower right corner. Turn the rotary knob to navigate the cursor around
the display. Push the knob to select highlighted options.
INDICATOR
DESCRIPTION
AC
The green AC Power LED will light to indicate that the monitor is connected to an AC power
source.
BATT
NOT USED
*NOTE! Qualified personnel may change the allowed alarm audio silence mode. See
Select the Alarm Audio
Silence Operation
in
Chapter 12: Service Menu
for details regarding alarm silence.
Figure 3.3: Keys