Chapter 3: Controls and Features
3-6
Veterinary 3 Parameter Advisor
®
Operation Manual
CONNECTOR
DESCRIPTION
INSTRUCTION
x
AC Power Connector
and Fuse Holder
Plug the AC power cord into the AC power receptacle at the back of the
monitor. When the other end is plugged into a grounded, three-wire
hospital-grade outlet, the AC Power LED (
X
) will light. The monitor
automatically switches between 110V and 220V AC line voltage
sources.
WARNING! Do not plug the monitor into an outlet controlled by a
wall switch.
The monitor’s fuse holder is below the AC power receptacle. The fuses
(5x20 mm fast-acting 1.6A/250V) are installed in the fuse holder.
y
Voltage Rating
The monitor’s AC rating is shown. If the AC rating does not match
the nominal voltage at your installation site, do not use the monitor.
Contact your authorized repair center for help.
Back Panel
CONNECTOR
DESCRIPTION
INSTRUCTION
u
Air Vents
The monitor has air vents at the top of the back panel and on the
bottom of the monitor.