Chapter 9: Non-invasive Blood Pressure
9-4
Veterinary 3 Parameter Advisor
®
Operation Manual
Figure 9.2: Attach the Cuff
WARNING! Verify cuff size is correct for the selected patient mode on the monitor.
Attach the cuff to the patient.
2.
Squeeze all the air out of the cuff.
a.
Place the cuff on a limb at the same level as the heart. The SurgiVet
b.
®
logo should be facing up, away from
the patient. The width of the cuff should be approximately 30-60% of the circumference of the limb. You
do not need to align the cuff along an artery.
Wrap the cuff around the limb and secure the Hook & Loop closure.
c.
Connect the NIBP supply hose to the monitor.
3.
WARNING! Make sure that hoses are not kinked, compressed, or restricted.
Figure 9.3: Connect the NIBP Supply Hose
Connect the cuff to the supply hose.
4.
The NIBP parameter box is always displayed; measured values will appear as soon as a blood pressure
•
measurement is taken.
ECG
NIBP
BP2
BP1
TEMP
SPO2
w
2
1
2
100-240AC
0.8A 50-60 Hz
F 1.6A/250V
REPLACE FUSE AS MARKED
RISK OF FIRE
WARNING:
CAUTION
Connect the NIBP