When the H-1000 is plugged in and turned ON and the disposable set is correctly
installed, a green"OPERATING" light will illuminate and the water temperature will begin
to increase on the display panel. If the disposable set is incorrectly installed and the unit is
turned ON, the yellow "CHECK DISPOSABLE" light will illuminate and an intermittent
alarm will sound.
Priming Disposable Sets
! WARNING !
Set up, priming and use require aseptic technique. Follow applicable hospital policy and
procedures.
! WARNING !
Remove all air from fluid bags before spiking with D-series disposable set.
! WARNING !
All air must be removed from fluid lines prior to patient connection. Failure to do so could
result in the introduction of air to the patient.
! WARNING !
Monitor fluid lines to ensure they are air-free. Never infuse fluids if there are bubbles in
the line between the filter chamber and patient connection.
1. CLOSE the Y-Set clamps.
2. Remove one bag spike cap and spike it into an air-free fluid bag with a twisting motion.
3. Hang the fluid bag from the adjustable I.V. pole and open the lead clamp to the I.V.
bag.
4. Squeeze the drip chamber to the I.V. bag so it is 1/3 to 1/2 full.
5. Remove the male luer cap on the patient line and open the Y-set clamps to the I.V.
bag.
6. Close the patient line clamp when the disposable set is completely primed. The filter
with gas vent will self prime.
7. Gently tap the filter with gas vent against the cabinet to release all trapped air. Priming
is complete. Make patient connection without entrapping air.
8. Adjust flow rate using roller clamp.
Filter with Gas Vent Replacement in I.V. Disposable Sets
Replace the filter with gas vent every THREE HOURS or when the filter becomes clogged
or air is slowly vented. Replace with Level 1
®
product number F-10 or F-30. Refer to
disposable packaging for appropriate filter replacement product number.
13