HOT LINE
®
Blood and Fluid Warmer
Technical Service Pack
46
S E C T I O N 5
M a i n t e n a n c e
Maintenance
Before each use
Carefully inspect the HOTLINE
®
for signs of damage, cracked or
split case, insecure poleclamp, damaged mains cable, etc.
If the Disposable Set does not install easily, lubricate the O-
Ring Seals.
Lubricating the O-Ring Seals
Using a cotton swab, apply a small amount of silicone grease
(Smiths Medical Part Number EZL 80-04-002) to the O-Rings in the
Disposable Set connector block.
If this does not solve the problem, reject the unit.
After each use
Wipe all external surfaces of the HOTLINE
®
with a soft cloth,
using an aqueous solution of mild detergent. If necessary, the
external surfaces may be disinfected using a solution of 10% bleach
in distilled water.
■
Do not autoclave.
■
Do not use alcohol or solvents.
■
Do not use abrasive cleaning agents.
■
Do not use cold sterilants.
■
Do not immerse any part of the HOTLINE
®
in liquids.