504-2038NUS
13
March 2005
3. Check the condition of the babyPAC 100 (continued)
xxviii)
Remove the existing battery and check when battery was last changed. If 12 months or
more has elapsed, insert a new battery of the correct type (part no. W269-023). If service
interval is less than 12 months, mark the battery, or service log for the ventilator, with the date
of replacement.
xxix) Replace any damaged labels.
xxx) Check that any accessories or spares added are MR Compatible.
Figure 1: Controls and fittings of the babyPAC 100 ventilator
Supply gas failure
visual alarm
Oxygen
concentration
PEEP/CPAP
Function Selector switch
Expiratory time T
EXP
Inspiratory time T
INS
High
inflation
pressure
alarm
Connector
Y-piece
Patient Inspiratory