WEIGHING WITH ELECTRONIC SCALE (optional)
❑
Press the
“
ON/ZERO
”
button to turn on the scale. If
“
LOW BAT
”
appears, replace 9V
battery located on the back of the scale. The scale will take about 5 seconds to zero itself.
❑
Attach the desired VOLARO sling, Fleece Cover, Recumbent Bars or anything else you
plan to use on the person being weighed. Make sure it is not touching anything.
❑
Press the
“
ON/ZERO
”
button –
The scale counts down for 3 seconds (DO NOT touch lift
during countdown or countdown will reset for additional 3 seconds.) This will place the
weight of the sling into memory and the scale will automatically deduct the weight of the
sling from the reading.
NOTE:
To switch from
Lbs.
to
Kilos
-
Press and hold the
LB
/
KG
button until
”
CONV
”
disappears. Press and hold again to convert back to
Lbs
. (Note: this may trig-
ger a need to recalibrate after weighing.)
❑
Now with the scale still on, remove the sling and place it beneath the person to be weighed.
NOTE
the appearance of the screen. You will see a flashing #. This is the weight of the
items previously weighed. They will automatically be subtracted from the person
’
s
weight. This reading will stay in memory even if the scale is turned off or the battery is
removed.
❑
Attach and lift the person clear of chair or bed. Once the reading is steady, the scale will
lock onto the reading to make it easier to reading.
NOTE:
Before you record weight,
Be sure that you are not pulling on the hanger
.
This will show a false weight. Let weight steady and record reading.
❑
Scale shuts off automatically after
one minute.
Press the recall button to review the last
recorded weight. (See Calibration Procedures on the next page.)
VOLARO CHARGER SPECIFICATIONS
Bulk Charge Current
1.5 Amps
Bulk Charge Voltage Range
10
-
14 Volts
Absorption Mode Current
1.5
-
.5 Amps
Absorption Mode Voltage Range
14.0
-
15.0 Volts
Float Voltage
13.6
-
13.8 Volts
AC Voltage In
115
-
130 VAC
Power In
30 Watts
13