18
Note: Year selection will scroll up from 2013. Month selection will scroll from 1 to 12. Day selection
will scroll from 1 to 31.
Note: Hour selection will scroll through 0 to 23. Minute and Second selections will scroll through 0 to
59.
Note: If batteries are removed from wrench for longer than 20 minutes, clock will revert to default
settings and must be re-entered at power on.
Setting Calibration Interval
This function will allow user to set calibration interval for when "CAL NEEDED" message will be
displayed.
1.
From Configure menu, use
UP
/DOWN
buttons to highlight
SET CAL INTRVAL
selection then
press
ENTER
button.
2.
CAL INTERVAL screen is displayed.
3.
Use
UP
/DOWN
buttons to change calibration interval.
Selectable Intervals:
DISABLED (factory default)
3 MON
6 MON
12 MON
4.
Press
ENTER
button to accept selection and exit to Configure menu.
Note: Clock Date and Time must be set before calibration interval will function. If batteries are
removed from wrench for longer than 20 minutes, clock will revert to default settings and must
be re-entered at power on.
ENTER
button
CAL INTRVAL
DISABLED
EXIT
MODE SETUP
CALIBRATION
SET DATE/TIME
SET CAL INTRVAL
ENTER
button
UP/DOWN
buttons
CAL INTRVAL
12 MON
EXIT
SET DATE
2013
/01/01
YR MON DAY
SET DATE
2013/
01
/01
YR MON DAY
SET DATE
2013/01/
01
YR MON DAY
EXIT
MODE SETUP
CALIBRATION
SET DATE/TIME
ENTER
button
UP/DOWN
buttons
ENTER
button
UP/DOWN
buttons
ENTER
button
UP/DOWN
buttons
SET TIME
12
: 00 : 00
HR MIN SEC
SET TIME
12 :
00
: 00
HR MIN SEC
SET TIME
12 : 00 :
00
HR MIN SEC
UP/DOWN
buttons
ENTER
button
UP/DOWN
buttons
ENTER
button
UP/DOWN
buttons
ENTER
button
ENTER
button
Summary of Contents for TECHANGLE
Page 119: ...118 ...