5.3 Cleaning
Always disconnect the scale from the power supply before cleaning.
Clean the scale as required using normal household cleaning agents. Ensure that no liquid enters the
scale. Wipe with a dry, soft cloth.
Never use abrasive or aggressive cleaning agents for cleaning. These agents can damage the scale.
5.6 Maintenance and service
The scale must be regularly subjected to a preventive inspection. To do this, load the scale with a
known weight and compare the value of the terminal with the test weight. In case of deviations,
maintenance or adjustment must be carried out.
The scale may only be opened and repaired by trained service personnel authorised by Soehnle
Industrial Solutions. If the scale does not work as intended, damage is suspected. In this case, the scale
must be taken to a service station authorised by Soehnle Industrial Solutions. In case of repair by an
authorised service station, only original spare parts may be used.
5.4 Disinfection
The 7835 personal scales may be disinfected with the following agents:
f
Denatured alcohol
f
Isopropyl (alcohol)
f
2% kohrsoline
f
1% aqueous solution of Socrena
f
5% Sagrotan
f
5 % Gigasept
The specific instructions for use must be observed.
5.5 Sterilisation
The 7835 stand scales must not be subjected to any sterilisation process.