2.6 Ten times resolution X10
The ten times resolution appears by pressing the
-key and the Zeroing key
simultaneously.
On non-approved scales, the high resolution appears as long as you press
the
-key and the Zeroing key.
2.7
Switching over the scales (measuring points)
The Basic Terminal 3010 can operate 2 analogue measuring points.
Toggle between the measuring points by pressing the INFO key followed by
the
-key (in succession).
2.8
Alibi memory
The alibi memory is activated in Setting mode (see separate description
470.702.078 User Mode). When the alibi memory is activated, an arrow points
to the corresponding symbol.
The arrow flashes as soon as the entered limit for full indication is reached.
Press the
-key or data request to accept a value to the alibi memory. To
do this, the print template or the data record must be configured accordingly
(see separate description 470.508.059 Data Interface).
Send a request to the alibi memory (view/print entries) in Setting mode:
Press
- and
-keys for 5 seconds until
UCAL I
appears.
Press
-key repeatedly until
UCAL 5
appears.
Press the
-key to confirm.
Then press the
-key until the Setting step 03 appears.
Press the
-key to confirm.
The number of the last data record saved appears. Call each data record by
changing the consecutive number using the arrow keys (
/
) and
display by confirming with the
-key.