7. ERROR MESSAGES
Error message
Cause /Remedy / Info
4
Ref weight too small
Scale unsuitable for part weight, if necessary change settings in
setting mode
5
Not possible to zero
Deviation from zero too large
6
Not possible to tare
Empty scale, scale in underload or overload, manual tare input
with tare value above maximum load
7
Not authorised to print
Selected function does not allow printing, changing or extending
function.
8
No unit switchover
Shift key locked or disabled, no 2nd unit defined.
11
No measuring point signal
Cable interruption or short circuit to load cell, measuring point
defective
23
Print error
Printer not plugged in or Off-Line, cable connection not available
or defective
50
Value below zero range limit
Measured value less than the stored zero point, scale
mechanically damaged, apply preload, readjust, replace load
cell.
51
Value above zero range limit
Measured value greater than stored zero point, scale
mechanically damaged, remove preload, readjust, replace load
cell.
52
Overload
Measured value across weighing range, unload scale, defective
load cell
53
Underload
Turn on scale with preload and then remove preload, switch off
display on the keypad, restore weighing platform to original
condition, switch on display again.
75
Replace backup battery
Replace battery
8. ADDITIONAL INFORMATION
Information concerning:
> User program I/O port programming (dynamic switching)
> Communication (Alibi memory, Interfaces, Print templates, Barcode, Batch processing, IO control)
> Print
> Service (Error messages, Software updates)
> and much more
can be found under:
https://www.soehnle-professional.com/en/site/documents
Please go to the customer center on our website www.soehnle-professional.com and select the terminal
3025 under Downloads/Industrial scales/Terminals.