26
Zero range at power-on
Query
off or on
When the scale is turned on, it sets the zero range of the weighing range to zero
when the power is turned on.
Selection only possible in the service area, subject to calibration.
Auto tare
Selection
off or on
off: No auto tare
on: Automatically tares the first weight value above the empty signal after stability
on the scale.
Factory setting: off
Additive tare
Selection
off or on
off: New tare overwrites old tare
one: New taring adds new tare to the existing tare.
Please note: The function Tare totalising on has an influence on tare sequences
and totals. The description in the manual always refers to the Total tare setting.
Factory setting: off
Second unit
Selection
> g
> t
> lb
> mg
> oz.
Can be activated as auxiliary display function in the display via key F3 Auxiliary
display in weighing mode.
In the selection, the adjusted unit is not displayed as the first unit.
Empty message
Input
of the percentage of the weighing range in which the empty message is displayed.
Enter the percentage using the numeric keypad.
(Info: Empty message is not shown on the display.)
Factory setting: 0.10 %.
Check mode
Query or input
off or on
on switches on a 10-fold higher resolution in the display constantly and
deactivates the x10 key.
Only possible with non-verifiable scales.
Serial number MP
Query
of serial number Measuring point (MP)
Software version MP
Query
of the software status of the measuring point
(MP)