Tare weighing and add-on weighing
Place the container on the base. Press the
key.
0
is displayed.
Place the weighing goods on the scale.
Add-on weighing:
Press the
key again until
0
is displayed. Add
the additional weighing goods. You can keep doing this up to the
maximum load weight.
Check weighing
Place the reference object on the scale. Tare. Remove the reference
object. Place the object under test on the scale and read off the
plus/minus difference on the display.
Total weight display
When objects are on the scale, press the
key.
0
is displayed.
Remove the goods for weighing. The total weight appears with a
minus sign.
Switching off
When the display shows
0
, press the
key. When operating the
scale with batteries or rechargeable batteries, the auto-off function
is active. When the display shows
0
, the scale switches off automati-
cally after 2 minutes.
Clear tare
When the scale is empty, press the
key again.
User calibration
Turn on the scale
, press and hold
key for one second then
press
key for three times, to enter calibration mode. At this
moment, LCD display will show its Zero Counts.
Press
Key to zero the counts showed.
Then load the calibration weight according to its capacity (500 g
or 5 kg).
Press
key again to calibrate the scale.
Wait till “CAL” disappears. Unload the calibration weight and test the
scale.