19
>READ
>SITE: DEFAULT
(List of Sensors in
DEFAULT site)
>DEFAULT_VW
SITE: ABC_STN
12.651 psi
READING SAVED
Notes:
1) VW readings displayed on four screens
2) First line of first screen is reading in engineering units
3) Other screens are readings in raw units
Notes:
1) List is sorted by SENSOR name
2) DEFAULT site has only one SENSOR
“DEFAULT_VW”
3) Other sites may have more sensors
Notes:
1) List is sorted by SITE name
2) DEFAULT site is always on the top of list
Save
PRESS -->
TEMP = 23.5 °C
FRQ = 2654.221 Hz
TEMP = 23.5 °C
Hz
2
/1K = 7044.965
TEMP = 23.5 °C
PER = 376.76 uS
TEMP = 23.5 °C
SITE: BCC_ST_STN
SITE: XYZ_QUARRY
SITE: XYZ_STN
.........................
Sensor ‘
DEFAULT_VW
’ uses the following default parameters of reading
and displaying:
•
Site name: DEFAULT
•
Sensor name: DEFAULT_VW
•
Sweep frequency: 1400-3500 Hz
•
Excitation V = 5V
•
Display Resolution = 0.1
•
Unit (raw) = Hz
2
/1K
•
Unit (eng) = Hz
2
/1K
•
R0=0
•
G=-1
•
Calibration=linear
•
Unit (temp)=
o
C.
In most cases, the ‘
DEFAULT_VW
’ sensor will allow you to take readings
from most type of VW sensors before you have the chance to define a
sensor list for the VWnote.
TIP